The FlexAbility™ SE catheter has a laser-cut 8Fr 4-mm flexible tip irrigated through laser-cut kerfs with a thermocouple 0.3 mm from the distal end. The TactiCath™ SE catheter has an 8Fr 3.5-mm tip and 6-irrigation port with a thermocouple 2.67 mm proximal to the tip. We investigated the impact of these differences on the efficacy and safety of radiofrequency (RF) applications.
RF applications at a range of powers (20 W, 30 W, and 40 W), contact forces (5 g, 15 g, and 25 g), and durations (10-60 s) using perpendicular/parallel catheter orientation were performed in excised porcine hearts. Lesion characteristics and incidence of steam pops were compared.
A total of 540 lesions were examined. The FlexAbility™ SE catheter produced smaller lesion depths (4.0 mm vs. 4.4 mm, p = 0.014 at 20 W; 4.6 mm vs. 5.6 mm, p = 0.015 at 30 W), surface areas (22.7mm vs. 29.2mm at 20 W, p = 0.005; 23.2mm vs. 28.7mm, p = 0.009 at 30 W), and volumes (126.1mm vs. 175.1mm, p = 0.018 at 20 W; 183.2mm vs. 304.3mm, p = 0.002 at 30 W) with perpendicular catheter placement. However, no differences were observed with parallel catheter placement. Steam-pops were significantly less frequently observed with the FlexAbility™ SE catheter (4% vs. 22%, p < 0.001) irrespective of catheter direction to the tissue. Multivariate analysis showed that use of the TactiCath™ SE catheter, power ≥ 40 W, contact force ≥ 25 g, RF duration > 30 s, parallel angle, and impedance drop ≥ 20Ω were significantly associated with occurrence of steam-pops.
The FlexAbility™ SE catheter reduced the risk of steam-pops but produced smaller lesions with perpendicular catheter placement compared to the TactiCath™ SE catheter.
FlexAbility™ SE 导管具有激光切割的 8Fr 4mm 柔性尖端,通过激光切割狭缝进行冲洗,其末端有一个距末端 0.3mm 的热电偶。TactiCath™ SE 导管具有 8Fr 3.5mm 的尖端和 6 个带有热电偶的冲洗口,距离尖端 2.67mm。我们研究了这些差异对射频 (RF) 应用效果和安全性的影响。
在离体猪心内,使用垂直/平行导管方向,在不同功率(20W、30W 和 40W)、接触力(5g、15g 和 25g)和时间(10-60s)下进行 RF 应用。比较了病变特征和蒸汽弹出的发生率。
共检查了 540 个病变。与 TactiCath™ SE 导管相比,FlexAbility™ SE 导管在垂直导管放置时产生的病变深度更小(20W 时为 4.0mm 对 4.4mm,p=0.014;30W 时为 4.6mm 对 5.6mm,p=0.015),表面积更小(20W 时为 22.7mm 对 29.2mm,p=0.005;30W 时为 23.2mm 对 28.7mm,p=0.009),体积更小(20W 时为 126.1mm 对 175.1mm,p=0.018;30W 时为 183.2mm 对 304.3mm,p=0.002)。然而,在平行导管放置时,没有观察到差异。无论导管相对于组织的方向如何,FlexAbility™ SE 导管的蒸汽弹出发生率明显较低(4%对 22%,p<0.001)。多变量分析显示,使用 TactiCath™ SE 导管、功率≥40W、接触力≥25g、RF 持续时间>30s、平行角度和阻抗下降≥20Ω与蒸汽弹出的发生显著相关。