Heart Center, Massachusetts General Hospital, Boston, Massachusetts.
Kansas University Medical Center, Kansas City, Kansas.
Heart Rhythm. 2017 Nov;14(11):1631-1636. doi: 10.1016/j.hrthm.2017.07.023. Epub 2017 Jul 20.
While the body of evidence showing the benefit of contact force (CF)-sensing catheters in improving atrial fibrillation ablation success rate is growing, real-world safety data of this technology are limited.
The aim of this study was to report the complication rate in patients undergoing ablation using a specific CF-sensing catheter (TactiCath Quartz CF ablation catheter, Abbott, St. Paul, MN) in a large worldwide database.
User-reported adverse event data from January 1, 2014 to December 31, 2015 were reviewed for the ablation catheter. These adverse events included cardiac perforation, cerebrovascular accident, atrioesophageal (AE) fistula, steam pop, and death. Rates of these adverse events, including geographic and temporal relationships, were calculated.
Adverse events were reported in 178 of 41,709 patients (adverse event rate 0.427%). Adverse events included 117 cardiac perforations (0.281%), 5 cerebrovascular accidents (0.012%), 10 AE fistulas (0.024%), 53 audible steam pops (0.127%), and 13 deaths (0.031%). The total event rate in European countries was 0.53% vs 0.38% in the United States (P = .0082). The AE fistula event rate peaked in late 2014 and early 2015 at 0.06% and decreased over time to a steady-state 0.01%, corresponding to the initial US release, subsequent increased use, and training on the manufacturer recommended settings.
Complication rates while using the TactiCath CF ablation catheter appear to decrease with training and experience. The major complication rate associated with this technology is low in the context of similar reports of conventional radiofrequency catheters.
尽管越来越多的证据表明接触力(CF)感应导管在提高房颤消融成功率方面具有优势,但该技术的实际安全性数据仍很有限。
本研究旨在报告在使用特定 CF 感应导管(雅培,明尼苏达州圣保罗市的 TactiCath Quartz CF 消融导管)的大型全球数据库中患者的并发症发生率。
回顾了 2014 年 1 月 1 日至 2015 年 12 月 31 日期间用于消融导管的用户报告的不良事件数据。这些不良事件包括心脏穿孔、脑血管意外、食-房(AE)瘘、蒸汽噗声和死亡。计算了这些不良事件的发生率,包括地理和时间关系。
在 41709 例患者中报告了 178 例不良事件(不良事件发生率为 0.427%)。不良事件包括 117 例心脏穿孔(0.281%)、5 例脑血管意外(0.012%)、10 例 AE 瘘(0.024%)、53 例可闻蒸汽噗声(0.127%)和 13 例死亡(0.031%)。欧洲国家的总事件发生率为 0.53%,而美国为 0.38%(P=0.0082)。AE 瘘事件发生率在 2014 年末和 2015 年初达到 0.06%的峰值,随着时间的推移逐渐降至稳定状态的 0.01%,与美国的首次发布、随后的使用增加以及制造商推荐设置的培训相对应。
在使用 TactiCath CF 消融导管时,并发症发生率似乎随着培训和经验的增加而降低。在类似的传统射频导管报告中,该技术的主要并发症发生率较低。