Department of Gastroenterology and Hepatology, Erasmus Medical Center, Rotterdam, The Netherlands.
Department of Medical Microbiology and Infectious Diseases, Erasmus Medical Center, Rotterdam, The Netherlands.
Endoscopy. 2022 Nov;54(11):1085-1090. doi: 10.1055/a-1814-4379. Epub 2022 May 5.
To mitigate duodenoscope contamination, recent design enhancements have primarily focused on the distal tip. However, the working channels remain unchanged, which may be linked to biofilm formation. We assessed the persistence of microorganisms, indicative of biofilm formation, in reprocessed duodenoscopes in a non-clinical endoscopic retrograde cholangiopancreatography (ERCP) simulation setting.
Three new duodenoscopes were over-soiled in non-clinical ERCP simulations followed by reprocessing. After 40 tests, the strain in the soil (Pa-type 1) was switched to a different strain (Pa-type 2) for 20 subsequent tests. Cultures of the tip and working channel were acquired after high level disinfection and overnight storage.
One duodenoscope showed persistent growth of from the fifth test until the end of the study. Pa-type 1 remained present until the end of the study in the cultures of this duodenoscope, even after discontinuation of exposure to that specific strain. The other two duodenoscopes only showed incidental contamination.
Persistent contamination by Pa-type 1 was seen in one out of three duodenoscopes after exposure to supraphysiological levels of gut microorganisms. No clear explanation was found for this persistent contamination as exposure and handling were identical and no abnormalities of this particular duodenoscope were identified by borescope inspection.
为了减轻十二指肠镜的污染,最近的设计改进主要集中在远端尖端。然而,工作通道保持不变,这可能与生物膜的形成有关。我们评估了在非临床内镜逆行胰胆管造影(ERCP)模拟环境中重新处理的十二指肠镜中微生物的持续存在,这些微生物表明生物膜的形成。
在非临床 ERCP 模拟中,对三个新的十二指肠镜进行过度污染,然后进行再处理。在 40 次测试后,土壤中的菌株(Pa 型 1)被切换为 20 次后续测试中的另一种菌株(Pa 型 2)。高水平消毒和过夜储存后,采集尖端和工作通道的培养物。
一台十二指肠镜从第五次测试开始一直到研究结束,都持续有 生长。Pa 型 1 直到研究结束时仍存在于该十二指肠镜的培养物中,即使在停止接触该特定菌株后也是如此。另外两个十二指肠镜仅显示偶然污染。
在暴露于超生理水平的肠道微生物后,三种十二指肠镜中的一种出现了 Pa 型 1 的持续污染。由于暴露和处理相同,并且没有通过内窥镜检查发现该特定十二指肠镜的异常,因此对这种持续污染没有明确的解释。