高频线阵探头定量超声评估人关节软骨早期骨关节炎。

Quantitative Ultrasound Assessment of Early Osteoarthritis in Human Articular Cartilage Using a High-Frequency Linear Array Transducer.

机构信息

Frederic L. Lizzi Center for Biomedical Engineering, Riverside Research, New York, New York, USA.

Berlin Institute of Health, Charité-Universitätsmedizin Berlin, Berlin, Germany.

出版信息

Ultrasound Med Biol. 2022 Aug;48(8):1429-1440. doi: 10.1016/j.ultrasmedbio.2022.03.006. Epub 2022 May 7.

Abstract

Quantitative ultrasound (QUS) assessment of osteoarthritis (OA) using high-frequency, research-grade single-element ultrasound systems has been reported. The objective of this ex vivo study was to assess the performance of QUS in detecting early OA using a high-frequency linear array transducer. Osteochondral plugs (n = 26) of human articular cartilage were scanned with ExactVu Micro-Ultrasound using an EV29L side-fire transducer. For comparison, the samples were also imaged with SAM200Ex, a custom 40-MHz scanning acoustic microscope with a single-element, focused transducer. Thirteen QUS parameters were derived from the ultrasound data. Magnetic resonance imaging (MRI) data, with T1 and T2 extracted as the quantitative parameters, were also acquired for comparison. Cartilage degeneration was graded from histology and correlated to all quantitative parameters. A maximum Spearman rank correlation coefficient (ρ) of 0.75 was achieved using a combination of ExactVu QUS parameters, while a maximum ρ of 0.62 was achieved using a combination of parameters from SAM200Ex. A maximum ρ of 0.75 was achieved using the T1 and T2 MRI parameters. This study illustrates the potential of a high-frequency linear array transducer to provide a convenient method for early OA screening with results comparable to those of research-grade single-element ultrasound and MRI.

摘要

已有研究报道了使用高频、研究级单阵元超声系统对骨关节炎(OA)进行定量超声(QUS)评估。本离体研究的目的是使用高频线性阵列换能器评估 QUS 在检测早期 OA 方面的性能。使用 ExactVu Micro-Ultrasound 及其 EV29L 侧发式换能器对 26 个人类关节软骨的骨软骨栓进行扫描。为了进行比较,还使用了具有单个聚焦换能器的定制 40MHz 扫描声学显微镜 SAM200Ex 对样本进行成像。从超声数据中得出了 13 个 QUS 参数。还获取了磁共振成像(MRI)数据,提取 T1 和 T2 作为定量参数进行比较。根据组织学对软骨退变进行分级,并与所有定量参数相关联。使用 ExactVu QUS 参数组合可获得最大的斯皮尔曼等级相关系数(ρ)为 0.75,而使用 SAM200Ex 参数组合可获得最大的 ρ 为 0.62。使用 T1 和 T2 MRI 参数可获得最大的 ρ 为 0.75。本研究表明,高频线性阵列换能器具有提供早期 OA 筛查便捷方法的潜力,其结果可与研究级单阵元超声和 MRI 相媲美。

相似文献

6
Minimally invasive ultrasound method for intra-articular diagnostics of cartilage degeneration.用于关节内软骨退变诊断的微创超声方法。
Ultrasound Med Biol. 2009 Sep;35(9):1546-54. doi: 10.1016/j.ultrasmedbio.2009.04.004. Epub 2009 Jun 26.
10

引用本文的文献

2
Reply.回复。
Arthritis Rheumatol. 2025 Apr 28. doi: 10.1002/art.43206.

本文引用的文献

2
Estimation of Cortical Bone Microstructure From Ultrasound Backscatter.从超声背散射估算皮质骨微观结构。
IEEE Trans Ultrason Ferroelectr Freq Control. 2021 Apr;68(4):1081-1095. doi: 10.1109/TUFFC.2020.3033050. Epub 2021 Mar 26.
7
Clinical utility and potential of ultrasound in osteoarthritis.超声在骨关节炎中的临床应用和潜力。
Radiol Med. 2019 Nov;124(11):1101-1111. doi: 10.1007/s11547-019-01013-z. Epub 2019 Mar 4.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验