Department of Urology, Osaka University Graduate School of Medicine.
Department of Medical Physics and Engineering, Division of Health Sciences, Osaka University Graduate School of Medicine.
Magn Reson Med Sci. 2023 Jul 1;22(3):345-351. doi: 10.2463/mrms.mp.2021-0125. Epub 2022 May 10.
This study aimed to investigate the ability of creatine-chemical exchange saturation transfer (Cr-CEST) technique assessed through 7-T MRI to evaluate cisplatin-induced testicular damage.
We used 8-10 weeks C57BL/6 mice (n = 10) that were divided into a control group (n = 5) and a cisplatin-treated group (n = 5). The cisplatin group received cisplatin at a dose of 15 mg/kg, via intraperitoneal injection, while the control group received saline. MR images of mouse testes were acquired under anesthesia 18 days after the injection using a horizontal 7-T scanner. The pulse sequence consisted of rapid acquisition with a relaxation enhancement (RARE) with magnetization transfer. The Z-spectra were collected using a 2000-ms saturation pulse at a B amplitude of 1.2 μT, with frequencies varying from -4.8 to +4.8 parts per million (ppm). Maps of magnetization transfer ratio with asymmetric analysis (MTR) were reconstructed at a Cr metabolite concentration of 1.8 ppm.
The Cr-CEST effect was significantly reduced in the cisplatin-treated group compared to the control group (MTR of control mice vs. cisplatin-treated mice: 6.9 [6-7.5] vs. 5.2 [4-5.5], P = 0.008). Correlation analysis revealed a strong correlation between the Cr-CEST effect and the pathological score (ρ = 0.93, P < 0.001).
Cr-CEST MRI can be useful for the evaluation of cisplatin-induced testicular damage in mice.
本研究旨在探讨通过 7-T MRI 评估肌酸化学交换饱和传递(Cr-CEST)技术评估顺铂诱导的睾丸损伤的能力。
我们使用 8-10 周龄 C57BL/6 小鼠(n=10),分为对照组(n=5)和顺铂处理组(n=5)。顺铂组腹腔注射顺铂 15mg/kg,对照组注射生理盐水。注射后 18 天,在麻醉下使用水平 7-T 扫描仪获取小鼠睾丸的 MR 图像。脉冲序列由具有弛豫增强的快速采集(RARE)和磁化转移组成。在 B 振幅为 1.2μT 的情况下,使用 2000ms 饱和脉冲采集 Z 谱,频率从-4.8 到+4.8 ppm。在 Cr 代谢物浓度为 1.8ppm 时,重建具有不对称分析的磁化转移比(MTR)图。
与对照组相比,顺铂处理组的 Cr-CEST 效应明显降低(对照组小鼠的 MTR 与顺铂处理组小鼠的 MTR 相比:6.9[6-7.5]比 5.2[4-5.5],P=0.008)。相关性分析显示 Cr-CEST 效应与病理评分之间存在很强的相关性(ρ=0.93,P<0.001)。
Cr-CEST MRI 可用于评估小鼠顺铂诱导的睾丸损伤。