Sci Rep. 2022 May 12;12(1):7915. doi: 10.1038/s41598-022-11914-1.
The purpose of this study was to compare the fixation stability of proximal fragments and the mechanical characteristics in proximal femur models of basicervical femoral neck fracture fixed by the femoral neck system (FNS) versus the dynamic hip screw. The mean axial stiffness was 234 ± 35 N/mm in the FNS group and 253 ± 42 N/mm in the DHS group, showing no significant difference (p = 0.654). Mean values for x-axis rotation, y-axis rotation, and z-axis rotation after cycle load were 2.2 ± 0.5°, 6.5 ± 1.5°, and 2.5 ± 0.6°, respectively, in the FNS group and 2.5 ± 0.7°, 5.8 ± 2.1°, and 2.2 ± 0.9°, respectively, in the DHS group, showing no significant differences (p = 0.324, p = 0.245, and p = 0.312, respectively). The mean values of cranial and axial migration of screws within the femoral head were 1.5 ± 0.3 and 2.1 ± 0.2 mm, respectively, in the FNS group and 1.2 ± 0.3 and 2.4 ± 0.3 mm, respectively, in the DHS group, showing no significant differences (p = 0.425 and p = 0.625, respectively). The average failure load at vertical load was 1342 ± 201 N in the FNS group and 1450 ± 196 N in the DHS group, showing no significant difference (p = 0.452). FNS fixation might provide biomechanical stability comparable to that of DHS for treating displaced basicervical femoral neck fractures in young adults.
本研究旨在比较股骨颈系统(FNS)与动力髋螺钉(DHS)固定颈干角型股骨颈骨折近端骨块固定的稳定性和股骨近端的力学特性。FNS 组的平均轴向刚度为 234±35N/mm,DHS 组为 253±42N/mm,差异无统计学意义(p=0.654)。FNS 组循环载荷后 x 轴旋转、y 轴旋转和 z 轴旋转的平均值分别为 2.2±0.5°、6.5±1.5°和 2.5±0.6°,DHS 组分别为 2.5±0.7°、5.8±2.1°和 2.2±0.9°,差异均无统计学意义(p=0.324、p=0.245 和 p=0.312)。FNS 组螺钉在股骨头内的头侧和轴向迁移的平均值分别为 1.5±0.3mm 和 2.1±0.2mm,DHS 组分别为 1.2±0.3mm 和 2.4±0.3mm,差异均无统计学意义(p=0.425 和 p=0.625)。FNS 组垂直加载的平均失效负荷为 1342±201N,DHS 组为 1450±196N,差异无统计学意义(p=0.452)。FNS 固定治疗年轻成人移位颈干角型股骨颈骨折可能提供与 DHS 相当的生物力学稳定性。