The use of contaminated water in high speed handpieces and ultrasonic handpieces presents potential hazards to both dentist and patient. This study has shown that after waterline sterilization, the use of a sterile, disposable membrane filter can eliminate the microflora from the water of an ultrasonic scaler for up to 48 hours. Similarly, the water from a high speed handpiece can remain bacteria free for up to 72 hours when a 0.45 micrometer pore size membrane filter is installed into the waterline. This system of decontamination by filtration may present a workable approach to the problem of contaminated water and merits further research and development. Commercial materials and equipment are identified in this report to specify the investigative procedures. Such identification does not simply recommendation or endorsement or that the materials and equipment are necessarily the best available for the purpose. Furthermore, the opinions expressed herein are those of the authors and are not to be construed as those of the Army Medical Department.

摘要

在高速手机和超声波手机中使用受污染的水会给牙医和患者带来潜在危害。本研究表明,在水线灭菌后,使用无菌一次性膜过滤器可使超声波洁牙机的水中微生物在长达48小时内消除。同样,当在水线中安装孔径为0.45微米的膜过滤器时,高速手机的水可在长达72小时内保持无菌。这种过滤净化系统可能是解决水污染问题的可行方法,值得进一步研究和开发。本报告中列出了商业材料和设备以说明调查程序。这种列举并非简单推荐或认可这些材料和设备必然是该用途的最佳选择。此外,本文所表达的观点仅为作者观点,不应被视为陆军医疗部的观点。