Al Hassan Mohammed A, Salem Omar A, Alabsi Emad J
Orthopedic Surgery, King Fahad Specialist Hospital, Dammam, SAU.
Orthopedics, King Fahad Specialist Hospital, Dammam, SAU.
Cureus. 2022 Apr 12;14(4):e24091. doi: 10.7759/cureus.24091. eCollection 2022 Apr.
Rotating hinged knee replacements are used to restore knee stability when intrinsic stability is lost in the form of soft tissue compression. With medical engineering advancements and improvements in arthroplasty, intrinsic stability can be achieved by an implant post system. We present the case of a 44-year-old female who presented with post-traumatic right knee multi-ligamentous instability and advanced secondary osteoarthritis following a traumatic knee dislocation two years ago. The patient initially underwent a hinged total knee replacement. After five years, she got dislocation of hinged total knee replacement that affected her condition and necessitated emergency admission for open reduction and revision. Most reported cases of rotating hinge prosthesis dislocation occurred during the first year of follow-up. However, our case dislocated after five years of follow-up due to dislodgement from the tibial tray with the polyethylene channel in the form of fatigue failure of the anti-dislocation mechanism.
当因软组织受压导致固有稳定性丧失时,旋转铰链式膝关节置换术用于恢复膝关节稳定性。随着医学工程的进步和关节成形术的改进,可通过植入后系统实现固有稳定性。我们报告了一例44岁女性病例,该患者两年前因膝关节创伤性脱位后出现创伤性右膝多韧带不稳定和晚期继发性骨关节炎。患者最初接受了铰链式全膝关节置换术。五年后,她的铰链式全膝关节置换术发生脱位,影响了她的病情,需要紧急入院进行切开复位和翻修。大多数报道的旋转铰链假体脱位病例发生在随访的第一年。然而,我们的病例在随访五年后因抗脱位机制疲劳失效,导致聚乙烯通道从胫骨托上脱落而发生脱位。