van den Besselaar A M, Hermans J, van der Velde E A, Bussemaker-Verduyn den Boer E, van Halem-Visser L P, Jansen-Grüter R, Loeliger E A
J Biol Stand. 1986 Oct;14(4):305-17. doi: 10.1016/0092-1157(86)90018-1.
A number of commercial rabbit tissue thromboplastins used in oral anticoagulant control have been calibrated against the first International Reference Preparation for thromboplastins. This was done in a three-stage procedure by one laboratory, each stage representing a different level of thromboplastin comparability. The calibration model recently recommended by ICTH and ICSH was tested. This model proved to be suitable, although a statistically significant aberration was observed for some of the thromboplastins. The bias introduced by using the model in these non-ideal cases was small compared to the overall variation of the International Normalized Ratio, being the universal scale for reporting the prothrombin time during oral anticoagulant control. Batch-to-batch calibration using lyophilized pooled plasmas could be reliably performed for several commercial thromboplastins.