IEEE Trans Med Imaging. 2022 Nov;41(11):3053-3061. doi: 10.1109/TMI.2022.3176801. Epub 2022 Oct 27.
X-ray photon-counting detectors capable of resolving the energies of single X-ray photons are critical in medical imaging, and a high count rate is essential for photon-counting detectors. Here, we report the performance of the perovskite CsPbBr3 single-crystal semiconductor detector for X-ray photon counting. The CsPbBr3 detector noise floor, energy response linearity, energy resolution, count rate, and polarization were evaluated. By fine-tuning the detector working conditions, our CsPbBr3 detector with a planar electrode can count an incident X-ray photon rate of ~0.099 and ~0.336 Mcps/pixel at 10% and 30% deadtime loss, respectively, with corresponding energy resolutions of ~18% at 59.5 keV and ~12% at 122 keV at same pulse processing conditions. We also demonstrated that our CsPbBr3 detectors show negligible polarization under an X-ray flux of ~0.45 M photons/s/mm2 for the typical timescale of multiple clinical X-ray scans, such as 1 s- 100 s. Our evaluation demonstrates the high potential of CsPbBr3 detectors for X-ray photon-counting applications in medical and industrial diagnostics with a lower cost than the current state-of-the-art.
X 射线光子计数探测器能够分辨单个 X 射线光子的能量,在医学成像中至关重要,而高计数率对于光子计数探测器是必不可少的。在这里,我们报告了钙钛矿 CsPbBr3 单晶半导体探测器在 X 射线光子计数方面的性能。评估了 CsPbBr3 探测器的噪声底、能量响应线性度、能量分辨率、计数率和极化。通过微调探测器工作条件,我们的具有平面电极的 CsPbBr3 探测器在 10%和 30%死时间损失下,分别可以以约 0.099 和 0.336 Mcps/pixel 的入射 X 射线光子率计数,在相同的脉冲处理条件下,相应的能量分辨率分别约为 59.5 keV 的 18%和约 122 keV 的 12%。我们还证明,在典型的多次临床 X 射线扫描的时间尺度下,例如 1 s-100 s,我们的 CsPbBr3 探测器在约 0.45 M 光子/s/mm2 的 X 射线通量下表现出可忽略的极化。我们的评估表明,CsPbBr3 探测器在医学和工业诊断中的 X 射线光子计数应用中具有很高的潜力,成本低于当前的最先进水平。