Department of Radiation Oncology, University of Nebraska Medical Center, Omaha, Nebraska, USA.
Department of Neurosurgery, University of Nebraska Medical Center, Omaha, Nebraska, USA.
World Neurosurg. 2022 Aug;164:e749-e754. doi: 10.1016/j.wneu.2022.05.043. Epub 2022 May 18.
We sought to compare the dosimetric accuracy of postoperative stereotactic body radiation therapy in a carbon-fiber (CF) versus titanium instrumented spine using a cadaveric model.
In situ cadaveric implantation of titanium and CF instrumentation and dosimeter chips in a thoracic spine. The cadaver underwent stereotactic body radiation therapy, and a dose of radiation therapy was calculated, measured, and compared. The sensors were placed in positions to provide data on dosimetry near the screws (within 1 cm) and between the screws. The differences between calculated and measured doses were reported as percentages.
There was a significant difference in the dosimetry from calculated versus measured values near the screws of CF compared with titanium (P = 0.0057) with a mean percentage difference of only 2.93 for CF and a much higher value of 19.32 for titanium near the screws. There was also greater variability in the percent difference for the 2 screw types, with differences ranging from -16.54% to 35.20% near titanium screws and -3.37% to 1.66% near CF screws.
More accurate dosimetry and radiation therapy delivery with CF screws compared with traditional titanium screws may have implications on optimal radiation delivery, as well as complication avoidance. This may be due to reduced scatter and thus lower variability in radiation delivery with the volumetric modulated arc therapy technique.
我们旨在通过尸体模型比较碳纤维(CF)与钛仪器化脊柱术后立体定向体放射治疗的剂量学准确性。
在胸段脊柱原位植入钛和 CF 器械及剂量计芯片。尸体接受立体定向体放射治疗,并计算、测量和比较放射治疗剂量。传感器放置在靠近螺钉(1cm 内)和螺钉之间的位置,以提供螺钉附近剂量学的数据。报告计算剂量与测量剂量之间的差异百分比。
与钛相比,CF 螺钉附近的计算剂量与测量剂量之间存在显著差异(P=0.0057),CF 的平均差异百分比仅为 2.93,而钛的差异百分比高达 19.32。2 种螺钉类型的百分比差异也更大,钛螺钉附近的差异范围为-16.54%至 35.20%,CF 螺钉附近的差异范围为-3.37%至 1.66%。
与传统钛螺钉相比,CF 螺钉具有更准确的剂量学和放射治疗传递,这可能对最佳放射治疗以及避免并发症有影响。这可能是由于容积调强弧形治疗技术降低了散射,从而降低了放射治疗的变异性。