Zite-Ferenczy F, Rüdel R
Pflugers Arch. 1978 Apr 25;374(1):97-100. doi: 10.1007/BF00585702.
A simple method is devised to record rapid sarcomere length changes of muscle fibres using a lateral effect diode. In the standard position the diffractometer records length changes between 1.65 and 3.8 micron, the output being linear 1 V/micron with a frequency response of -3 dB at 1.2 kHz. The absolute error is less than 0.05 micron between 1.65 and 2.80 micron and less than 0.1 micron between 2.81 and 3.30 micron. The resolution of length changes is less than 0.005 micron over the whole range. By varying the detector position the length range can be extended to either side, and spatial resolution can be improved at the expense of length range.
设计了一种简单的方法,使用侧向效应二极管记录肌肉纤维的快速肌节长度变化。在标准位置,衍射仪记录的长度变化范围为1.65至3.8微米,输出为线性,1伏/微米,在1.2千赫时频率响应为-3分贝。在1.65至2.80微米之间绝对误差小于0.05微米,在2.81至3.30微米之间小于0.1微米。在整个范围内,长度变化的分辨率小于0.005微米。通过改变探测器位置,长度范围可以向两侧扩展,并且可以以长度范围为代价提高空间分辨率。