Nilsson B W, Rikner G, Wolgast M
Scand J Clin Lab Invest. 1977 May;37(3):201-8. doi: 10.3109/00365517709091483.
The intravenous isotope method has been examined. The method uses 99Tc as the indicator with recording of the radioactivity with two gamma-dectors placed bilaterally over the skull and oriented over the bifurcation of the middle cerebral artery. 65% of the total curve obtained was found to be generated by activity passing in the ipsilateral internal carotid artery, 25% in the contralateral internal carotid artery and less than 10% by activity passing in the two external carotid arteries. The counting rate efficiency for an intracerebral location was 19.4 +/- 1.8 times the count rate from the same activity placed at 1 m distance in air which in turn was 5 cps/muCi. The variability in efficiency from patient to patient was thus estimated at +/- 9%. The duration of the input bolus had a marked influence on the vurve, which will be of importance since it showed a rather large variability from patient. Using the peak activity as an index of the cerebral blood flow the intravenous method showed a surprisingly good correlation with the 133Xe wash-out method; the correlation coefficient was 0.93 and the variability at slightly subnormal values estimated at +/- 15%. It is concluded that the method can be used in the case of limited demands on the accuracy of the cerebral blood flow determinations.
已对静脉注射同位素法进行了研究。该方法使用99Tc作为指示剂,通过双侧置于颅骨上方并对准大脑中动脉分叉处的两个γ探测器记录放射性。发现所获得的总曲线中65%由同侧颈内动脉中的活性产生,25%由对侧颈内动脉中的活性产生,而由两个颈外动脉中的活性产生的不到10%。脑内位置的计数率效率是在空气中1米距离处相同活性计数率的19.4±1.8倍,而后者又为5 cps/μCi。因此,患者之间效率的变异性估计为±9%。输入团注的持续时间对曲线有显著影响,这一点很重要,因为它在患者之间显示出相当大的变异性。以峰值活性作为脑血流量的指标,静脉注射法与133Xe洗脱法显示出惊人的良好相关性;相关系数为0.93,略低于正常的值的变异性估计为±15%。得出的结论是,在对脑血流量测定的准确性要求有限的情况下,可以使用该方法。