Kabiş Burak, Tutar Hakan, Gündüz Bülent, Aksoy Songül
Department of Audiology, Gazi University Faculty of Health Sciences, Ankara, Turkey.
Department of Otolaryngology Head and Neck Surgery, Gazi University Faculty of Medicine, Ankara, Turkey.
Turk Arch Otorhinolaryngol. 2022 Mar;60(1):16-22. doi: 10.4274/tao.2022.2022-1-4. Epub 2022 May 12.
Vestibulo-ocular reflex gain at 40, 60, and 80 ms following the head movement start is calculated as the instantaneous gain. The purpose of this study was to investigate the reliability of instantaneous gain values at 40, 60, and 80 ms with testing and retesting in healthy adults.
The study was conducted with Interacoustics EyeSeeCam vHIT (Interacoustics, Denmark), and 42 healthy adults were evaluated twice at half-hour intervals (test and retest) by the same practitioner. Agreement of mean gain, gain asymmetry, and instantaneous gain was evaluated using a paired samples t-test.
Mean age of the participants was 33.62±11.17; 38.1% were male and 61.9% were female. In the degree of the agreement, paired sample correlation (r) between test and retest results of the horizontal semicircular canals was found to be higher than those of the vertical semicircular canals. Moreover, the highest correlation between test and retest for instantaneous gain, calculated for only horizontal semicircular canals, was found at 80 ms on each side (0.791; 0.838, right and left, respectively), while the lowest correlation between these parameters was found between the gain asymmetry values.
The video head impulse test used in studies calculates the mean gain in approximately at 60 ms. However, the higher correlation between mean gain values at 80 ms in our findings indicates that gain calculation strategies and techniques for latencies should be discussed. Additionally, the low correlation of vertical semicircular canals for mean gain and gain asymmetry between semicircular canal pairs, which clearly shows that more standard and more reliable methods should be developed.
头部运动开始后40、60和80毫秒时的前庭眼反射增益被计算为瞬时增益。本研究的目的是通过对健康成年人进行测试和复测,来探究40、60和80毫秒时瞬时增益值的可靠性。
本研究使用丹麦Interacoustics公司的EyeSeeCam视频头脉冲测试(vHIT)设备,由同一名从业者对42名健康成年人每隔半小时进行两次评估(测试和复测)。使用配对样本t检验评估平均增益、增益不对称性和瞬时增益的一致性。
参与者的平均年龄为33.62±11.17岁;男性占38.1%,女性占61.9%。在一致性程度方面,水平半规管测试和复测结果之间的配对样本相关性(r)高于垂直半规管。此外,仅针对水平半规管计算的瞬时增益在测试和复测之间的最高相关性出现在每侧80毫秒时(分别为0.791和0.838,右侧和左侧),而这些参数之间的最低相关性出现在增益不对称值之间。
研究中使用的视频头脉冲测试大约在60毫秒时计算平均增益。然而,我们的研究结果显示80毫秒时平均增益值之间的相关性更高,这表明应讨论增益计算策略和潜伏期的技术。此外,垂直半规管在平均增益和半规管对之间增益不对称性方面的相关性较低,这清楚地表明应开发更标准、更可靠的方法。