有源植入式心血管设备在职业低频电场和磁场中的干扰阈值:数值和体外研究。

Interference thresholds for active implantable cardiovascular devices in occupational low-frequency electric and magnetic fields: a numerical and in vitro study.

机构信息

Université de Lorraine, Institut Jean Lamour (UMR 7198), CNRS, 2 allée André Guinier, Campus Artem, 54000 Nancy, France.

Université de Lorraine, Institut Jean Lamour (UMR 7198), CNRS, 2 allée André Guinier, Campus Artem, 54000 Nancy, France.

出版信息

Med Eng Phys. 2022 Jun;104:103799. doi: 10.1016/j.medengphy.2022.103799. Epub 2022 Apr 13.

Abstract

In light of concerns regarding the occupational safety and health of workers wearing active implantable medical devices (AIMDs), this study aims to investigate the potential risks of electromagnetic interference (EMI) between AIMDs and low-frequency 50/60 Hz electromagnetic fields (EMFs) in the workplace. A total of 58 AIMDs, consisting of pacemakers (PMs) and implantable cardiac defibrillators (ICDs) of different brands, models, and configurations were tested to determine the immunity thresholds for high-voltage electric fields (EFs) and magnetic fields (MFs) at 50/60 Hz. The EFs and MFs at the levels in workplaces are reproduced by setups using Helmholtz coils and aluminum plates, respectively, to ensure that the EM/MF exposures are controllable and reproducible. The EMI thresholds were recorded by observing the occurrences of PM or ICD dysfunctions. In addition, numerical studies on anatomical models were carried out using CST® software. The results indicate that the recorded thresholds all exceed the EF and MF public exposure limits given in the ICNIRP 2010 guidelines. No dysfunction was observed among four ICDs tested under MF exposure up to 2750 µT at 50 Hz and 2480 µT at 60 Hz. However, among the 43 PMs and 11 ICDs tested under EF exposures, potential hazards may occur below the occupational exposure level proposed in the ICNIRP guidelines.

摘要

鉴于佩戴主动植入式医疗器械(AIMD)的工作人员的职业安全和健康问题,本研究旨在探讨 AIMD 与工作场所低频 50/60Hz 电磁场(EMF)之间的电磁干扰(EMI)的潜在风险。共有 58 个 AIMD,包括不同品牌、型号和配置的起搏器(PM)和植入式心脏除颤器(ICD),用于测试在 50/60Hz 时高电压电场(EF)和磁场(MF)的免疫阈值。使用亥姆霍兹线圈和铝板分别设置工作场所中电场和磁场的水平,以确保 EM/MF 暴露可控制且可重复。通过观察 PM 或 ICD 功能障碍的发生来记录 EMI 阈值。此外,还使用 CST®软件对解剖模型进行了数值研究。结果表明,记录的阈值均超过了 ICNIRP 2010 指南中给出的 EF 和 MF 公众暴露限值。在 50Hz 时高达 2750μT,60Hz 时高达 2480μT 的磁场暴露下,对四个 ICD 进行测试时未观察到功能障碍。然而,在对 EF 暴露进行测试的 43 个 PM 和 11 个 ICD 中,在 ICNIRP 指南中提出的职业暴露水平以下可能会出现潜在危险。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验