紧凑型磁共振兼容测力计及其在运动应激心脏磁共振中的应用:初步研究。

Compact MR-compatible ergometer and its application in cardiac MR under exercise stress: A preliminary study.

机构信息

Department of Radiology, West China Hospital of Sichuan University, Chengdu, Sichuan, China.

Molecular Imaging Center, West China Hospital of Sichuan University, Chengdu, Sichuan, China.

出版信息

Magn Reson Med. 2022 Oct;88(4):1927-1936. doi: 10.1002/mrm.29311. Epub 2022 Jun 1.

Abstract

PURPOSE

To develop a compact MR-compatible ergometer for exercise stress and to initially evaluate the reproducibility of myocardial native T1 and myocardial blood flow (MBF) measurements during exercise stress performed on this ergometer.

METHODS

The compact ergometer consists of exercise, workload, and data processing components. The exercise stress can be achieved by pedaling on a pair of cylinders at a predefined frequency with adjustable resistances. Ten healthy subjects were recruited to perform cardiac MRI scans twice in a 3.0T MR scanner, at different days to assess reproducibility. Myocardial native T1 and MBF were acquired at rest and during a moderate exercise. The reproducibility of the two tests was determined by the intra-group correlation coefficient (ICC) and coefficient of variation (CoV).

RESULTS

The mean exercise intensity in this pilot study was 45 Watts (W), with an exercise duration of 5 min. Stress induced a significant increase in systolic blood pressure (from 113 ± 11 mmHg to 141 ± 12, P < 0.05) and maximal increase in heart rate by 74 ± 19%. The rate pressure product increased two-fold (P < 0.001). Excellent reproducibility was demonstrated in native T1 during the exercise (CoV = 3.0%), whereas the reproducibility of MBF and myocardial perfusion reserve during the exercise was also good (CoV = 10.7% and 8.8%, respectively).

CONCLUSION

This pilot study demonstrated that it is possible to acquire reproducible measurements of myocardial native T1 and MBF during the exercise stress in healthy volunteers using our new compact ergometer.

摘要

目的

开发一种用于运动应激的紧凑型磁共振兼容测力计,并初步评估在该测力计上进行运动应激时心肌固有 T1 和心肌血流 (MBF) 测量的可重复性。

方法

紧凑型测力计由运动、工作量和数据处理组件组成。运动应激可以通过在一对带有可调节阻力的圆柱体上以预定频率踩踏来实现。招募了 10 名健康受试者,在不同的日子里在 3.0T 磁共振扫描仪上进行两次心脏 MRI 扫描,以评估可重复性。在休息和中度运动期间采集心肌固有 T1 和 MBF。通过组内相关系数 (ICC) 和变异系数 (CoV) 确定两次测试的可重复性。

结果

在这项初步研究中,平均运动强度为 45 瓦特 (W),运动持续时间为 5 分钟。应激引起收缩压显著升高(从 113±11mmHg 增加到 141±12mmHg,P<0.05)和最大心率增加 74±19%。心率压力产物增加了两倍(P<0.001)。在运动过程中,固有 T1 的可重复性非常好(CoV=3.0%),而运动期间 MBF 和心肌灌注储备的可重复性也很好(CoV 分别为 10.7%和 8.8%)。

结论

这项初步研究表明,使用我们的新型紧凑型测力计,有可能在健康志愿者的运动应激期间获得可重复的心肌固有 T1 和 MBF 测量值。