Blood glucose monitoring by children at home: a comparison of methods.

作者信息

Anderson D G, Gleeson M, Boulton T J

出版信息

Aust Paediatr J. 1986 Nov;22(4):309-12. doi: 10.1111/j.1440-1754.1986.tb02155.x.

Abstract

The accuracy of measurements of blood glucose performed at home by the index children and/or parents was compared using the Boehringer Mannheim reagent strips, Ames Visidex-2 strips, an Ames Glucometer and a Boehringer Mannheim Reflolux reflectance meter. Capillary plasma samples, collected simultaneously, were later analysed by a Beckman glucose oxidase analyser. The coefficient of correlation between the self-monitored results (SMR) and the laboratory-analysed results (LAR) ranged from 0.59 to 0.92, the composite being 0.74. The mean difference between the SMR and the LAR was 1.0 mmol/l. The occurrence of potential errors of management due to incorrect SMR results was determined. On four occasions, the SMR was greater than 6 mmol/l when the LAR was less than 3 mmol/l, and on 36 occasions a discrepancy greater than 50% occurred between the two values, without resulting hypoglycaemia.

摘要

使用勃林格殷格翰试剂条、艾姆斯Visidex - 2试纸条、一台艾姆斯血糖仪和一台勃林格殷格翰Reflolux反射仪,对指标儿童和/或家长在家中进行的血糖测量准确性进行了比较。同时采集的毛细血管血浆样本随后由贝克曼葡萄糖氧化酶分析仪进行分析。自我监测结果(SMR)与实验室分析结果(LAR)之间的相关系数在0.59至0.92之间,综合为0.74。SMR与LAR之间的平均差异为1.0 mmol/l。确定了由于SMR结果不正确而导致潜在管理错误的发生情况。有4次,当LAR低于3 mmol/l时,SMR大于6 mmol/l,并且有36次两个值之间出现了大于50%的差异,但未导致低血糖。