Marshall S M, Alberti K G
Diabetes Care. 1983 Nov-Dec;6(6):543-7. doi: 10.2337/diacare.6.6.543.
We have compared the performance of the newly available blood glucose test strip, Visidex (Ames Division, Miles Laboratories Ltd., Stoke Poges, Slough, United Kingdom), designed for visual reading, with that of the already available BM-Test-Glycemie 20-800 [Boehringer Mannheim Corporation (London) Ltd., Lewes, East Sussex, United Kingdom]. The new test strip proved slightly less accurate over a wide range of blood glucose values and was less precise at normal and high glucose values. The color of the Visidex faded more rapidly on storage and errors in timing of the blood strip reaction caused greater inaccuracies. At present BM-Test-Glycemie 20-800 remains the visual glucose test strip of choice, although even with this strip there was a wide scatter of values when used by patients.
我们比较了新推出的用于目测读数的血糖测试条Visidex(迈尔斯实验室有限公司艾姆斯分部,英国联合王国伯克郡斯托克波吉斯)与已有的BM-Test-Glycemie 20-800(德国勃林格殷格翰公司(伦敦)有限公司,英国东萨塞克斯郡刘易斯)的性能。新测试条在较宽的血糖值范围内被证明准确性略低,在正常和高血糖值时精密度也较低。Visidex测试条的颜色在储存时褪色更快,血条反应计时误差导致的不准确程度更高。目前,BM-Test-Glycemie 20-800仍然是首选的目测血糖测试条,不过即便使用这款测试条,患者使用时测得的值也有很大离散度。