Blood Press Monit. 2022 Oct 1;27(5):353-356. doi: 10.1097/MBP.0000000000000608. Epub 2022 May 27.
The aim of this study was to evaluate the accuracy of the watch-type automated oscillometric wrist blood pressure (BP) monitor HUAWEI WATCH D in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard.
Subjects were recruited to fulfill the age, sex, BP and cuff distributions of the AAMI/ESH/ISO Universal Standard in the general population using the same arm sequential BP measurement method. Two cuffs of the test device were used for wrist circumference of 13-16 cm (medium) and 16-20 cm (large), respectively.
One-hundred and nine subjects were recruited and 85 were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was -1.4 ± 6.47/-0.2 ± 5.85 mmHg (systolic/diastolic). For criterion 2, the SD of the average BP differences between the test device and reference BP per subject was 5.66/5.48 mmHg (systolic/diastolic).
The HUAWEI WATCH D watch-type wrist BP monitor fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard in the general population and can be recommended for self-measurement.
本研究旨在根据医疗器械促进协会/欧洲高血压学会/国际标准化组织(AAMI/ESH/ISO)通用标准评估华为 WATCH D 腕式自动示波法血压(BP)监测仪在普通人群中的准确性。
采用相同手臂顺序 BP 测量方法,根据 AAMI/ESH/ISO 通用标准在普通人群中的年龄、性别、BP 和袖带分布招募受试者。该测试设备的两个袖带分别适用于 13-16cm(中号)和 16-20cm(大号)的腕周。
共招募 109 名受试者,其中 85 名进行了分析。对于验证标准 1,测试设备和参考 BP 读数之间差值的平均值±SD 为-1.4±6.47/-0.2±5.85mmHg(收缩压/舒张压)。对于标准 2,每个受试者的测试设备和参考 BP 之间的平均 BP 差值的 SD 为 5.66/5.48mmHg(收缩压/舒张压)。
华为 WATCH D 腕式腕部 BP 监测仪满足普通人群中 AAMI/ESH/ISO 通用标准的所有要求,可推荐用于自我测量。