Clinic for Nuclear Medicine and Competence Centre for Thyroid Diseases, Ente Ospedaliero Cantonale, Bellinzona, and University Hospital and University of Zurich, Zurich, Switzerland.
Evgenideion Hospital, Unit of Endocrinology, Diabetes and Metabolism, University of Athens, Athens, Greece.
Endocrine. 2022 Aug;77(2):333-339. doi: 10.1007/s12020-022-03098-5. Epub 2022 Jun 11.
Measurement of thyroid-stimulating hormone (TSH) and free thyroxine (FT4) is important for assessing thyroid dysfunction. After changing assay manufacturer, high FT4 versus TSH levels were reported at Ente Ospedaliero Cantonale (EOC; Bellinzona, Switzerland).
Exploratory analysis used existing TSH and FT4 measurements taken at EOC during routine clinical practice (February 2018-April 2020) using Elecsys® TSH and Elecsys FT4 III immunoassays on cobas® 6000 and cobas 8000 analyzers (Roche Diagnostics). Reference intervals (RIs) were estimated using both direct and indirect (refineR algorithm) methods.
In samples with normal TSH levels, 90.9% of FT4 measurements were within the normal range provided by Roche (12-22 pmol/L). For FT4 measurements, confidence intervals (CIs) for the lower end of the RI obtained using direct and indirect methods were lower than estimated values in the method sheet; the estimated value of the upper end of the RI (UEoRI) in the method sheet was within the CI for the UEoRI using the direct method but not the indirect method. CIs for the direct and indirect methods overlapped at both ends of the RI. The most common cause of increased FT4 with normal TSH was identified in a subset of patients as use of thyroxine therapy (72.6%).
It is important to verify RIs for FT4 in the laboratory population when changing testing platforms; indirect methods may constitute a convenient tool for this. Applying specific RIs for selected subpopulations should be considered to avoid misinterpretations and inappropriate clinical actions.
测定促甲状腺激素(TSH)和游离甲状腺素(FT4)对于评估甲状腺功能障碍非常重要。在更换检测试剂生产商后,瑞士洛桑州立医院(Ente Ospedaliero Cantonale,EOC)报告了 FT4 升高而 TSH 正常的情况。
使用 Roche Elecsys TSH 和 Elecsys FT4 III 免疫分析法,在 cobas® 6000 和 cobas 8000 分析仪上(罗氏诊断公司),对 EOC 于 2018 年 2 月至 2020 年 4 月间常规临床检测中获取的 TSH 和 FT4 数据进行探索性分析。采用直接法和间接法(refineR 算法)分别对参考区间(RI)进行了估计。
在 TSH 正常的样本中,90.9%的 FT4 测量值处于罗氏提供的正常范围内(12-22 pmol/L)。对于 FT4 测量值,直接法和间接法获得的 RI 下限的置信区间(CI)低于方法表中估计值;方法表中 RI 上限(UEoRI)的估计值在直接法的 CI 内,但不在间接法的 CI 内。RI 两端的直接法和间接法 CI 存在重叠。在接受甲状腺素治疗的患者亚组中,发现了最常见的 FT4 升高而 TSH 正常的原因(72.6%)。
在更换检测平台时,实验室人群中需要验证 FT4 的 RI;间接法可能是一种方便的工具。对于选定的亚人群,应考虑应用特定的 RI,以避免错误解读和不适当的临床措施。