Santos B A, Pastora R, DelMonte M A, O'Donnell F E
J Cataract Refract Surg. 1987 Mar;13(2):127-30. doi: 10.1016/s0886-3350(87)80125-6.
We performed two tissue culture experiments designed to compare the effects of various posterior chamber optics on lens epithelium. In the first experiment, we recorded, by phase contrast microphotography, the migration of rabbit lens epithelium exoplants placed adjacent to the optic of various posterior chamber lenses. In the second experiment, we used phase contrast microphotographs to document the effects of various posterior chamber optics when gently placed on a confluent layer of rabbit lens epithelium. From our in vitro studies, we conclude the following: There is inhibition of lens epithelial migration and even cytotoxic effects from direct contact with polymethylmethacrylate optics; glass optics have appreciably less effect on lens epithelium; polymethylmethacrylate optics with ridges (complete annulus or incomplete) do not inhibit lens epithelial migration as well as planoconvex lenses, and they do not have a cytotoxic effect except at the points of contact between the ridge and the supporting surface.
我们进行了两项组织培养实验,旨在比较各种后房型光学部件对晶状体上皮的影响。在第一个实验中,我们通过相差显微镜摄影记录了置于各种后房型晶状体光学部件旁的兔晶状体上皮外植体的迁移情况。在第二个实验中,我们使用相差显微镜照片记录了各种后房型光学部件轻轻置于兔晶状体上皮汇合层时的影响。从我们的体外研究中,我们得出以下结论:与聚甲基丙烯酸甲酯光学部件直接接触会抑制晶状体上皮迁移,甚至产生细胞毒性作用;玻璃光学部件对晶状体上皮的影响明显较小;有脊(完整环或不完整)的聚甲基丙烯酸甲酯光学部件对晶状体上皮迁移的抑制作用不如平凸透镜,并且除了脊与支撑表面的接触点外,它们没有细胞毒性作用。