Chiang H, Zeidman E J, Alarcon A, Raz S
J Urol. 1987 May;137(5):936-8. doi: 10.1016/s0022-5347(17)44301-1.
Solid state microtransducers are used widely in urodynamics to record bladder and intraluminal urethral pressures. They are sensitive and capable of giving high quality, reproducible pressure recordings. They do not require the determination of a reference point but they are expensive and fragile, and calibration may be difficult to perform. We describe our experience with a new catheter for pressure recording based on fiberoptic technology. In addition to the advantages of the conventional microtipped catheters, the new catheter is thinner (4F), easier to calibrate and less expensive. Controlled testing of the new catheter was performed in an experimental setting and in clinical trials during urodynamic evaluation. The catheter provides reliable pressure recordings compared to standard microtipped catheters.
固态微传感器在尿动力学中被广泛用于记录膀胱和尿道腔内压力。它们灵敏度高,能够提供高质量、可重复的压力记录。它们不需要确定参考点,但价格昂贵且易碎,校准可能难以进行。我们描述了我们使用基于光纤技术的新型压力记录导管的经验。除了传统微尖端导管的优点外,新型导管更细(4F),更容易校准且成本更低。在实验环境和尿动力学评估期间的临床试验中对新型导管进行了对照测试。与标准微尖端导管相比,该导管能提供可靠的压力记录。