Heiliger R, Geks J, Mittermayer C
J Biomed Eng. 1987 Apr;9(2):128-33. doi: 10.1016/0141-5425(87)90023-9.
Two different mechanical heart valves with annulus diameters 21-29 mm, (five Björk-Shiley monostrut tilting disc valves and five Duromedics bileaflet valves) have been tested in pulsatile flow in the mitral position of a mock circulation. Reflux, pressure, and orifice area have been measured while cardiac output was varied between 2 and 6 1 min-1. Insufficiency, mean orifice area, discharge coefficient, and performance and efficiency indices have been calculated. Mean values of insufficiency for the Björk-Shiley monostrut valves varied between 4.8 and 17.2% while the corresponding values for the Duromedics valves were in the range 6.1-17.3%. Mean values for orifice areas of the Björk-Shiley monostrut valves increased with the larger valve sizes from 101.1 to 210.2 mm2; for the Duromedics valves the area range was 134.5-262.9 mm2. Because of the larger orifice areas the values of discharge coefficient and performance index for the Duromedic valves were higher than those for the Björk-Shiley monostrut valves. As the insufficiency of the two mechanical valves was similar, and the orifice area of the bileaflet valves was greater than that of the tilting disc valves, Duromedics valves gave higher valves for the efficiency index, which varied between 0.31 and 0.39; for Björk-Shiley monostrut valves the index varied between 0.24 and 0.28 under the same test conditions. This hydrodynamic in vitro comparison of mechanical heart valves showed that the Duromedics bileaflet valves were superior to the Björk-Shiley tilting disc valves.
两种不同的机械心脏瓣膜,瓣环直径为21 - 29毫米(五个Björk - Shiley单支柱倾斜盘式瓣膜和五个Duromedics双叶瓣膜),已在模拟循环的二尖瓣位置进行了脉动流测试。在心脏输出量在2至6升/分钟之间变化时,测量了反流、压力和孔口面积。计算了反流不足、平均孔口面积、流量系数以及性能和效率指标。Björk - Shiley单支柱瓣膜的反流不足平均值在4.8%至17.2%之间变化,而Duromedics瓣膜的相应值在6.1% - 17.3%范围内。Björk - Shiley单支柱瓣膜的孔口面积平均值随着瓣膜尺寸增大从101.1平方毫米增加到210.2平方毫米;Duromedics瓣膜的面积范围是134.5 - 262.9平方毫米。由于孔口面积较大,Duromedics瓣膜的流量系数和性能指标值高于Björk - Shiley单支柱瓣膜。由于两种机械瓣膜的反流不足相似,且双叶瓣膜的孔口面积大于倾斜盘式瓣膜,Duromedics瓣膜的效率指标更高,在0.31至0.39之间变化;在相同测试条件下,Björk - Shiley单支柱瓣膜的该指标在0.24至0.28之间变化。这种机械心脏瓣膜的体外流体动力学比较表明,Duromedics双叶瓣膜优于Björk - Shiley倾斜盘式瓣膜。