Matthaei D, Haase A
Magn Reson Med. 1987 Mar;4(3):302-5. doi: 10.1002/mrm.1910040312.
A simple modification of FLASH (Fast Low Angle SHot) MR imaging, which results in a variable reduction of the intensity of flowing blood and flow artifacts in transaxial tomograms, is reported. Here a nonselective radiofrequency pulse of variable flip angle is used before the acquisition of each projection in FLASH imaging to saturate flowing blood within the whole volume (SYstemic Saturation = SYS-FLASH).
据报道,对快速低角度激发(FLASH)磁共振成像进行简单修改,可使横断面断层图像中流动血液的强度和流动伪影有不同程度的降低。在此方法中,在FLASH成像的每次投影采集之前,使用可变翻转角的非选择性射频脉冲使整个容积内的流动血液饱和(全身饱和=SYS-FLASH)。