Suppr超能文献

快速FLASH NMR图像中的血流抑制

Flow suppression in rapid FLASH NMR images.

作者信息

Frahm J, Merboldt K D, Hänicke W, Haase A

出版信息

Magn Reson Med. 1987 Apr;4(4):372-7. doi: 10.1002/mrm.1910040408.

Abstract

Rapid FLASH (fast low angle shot) NMR images are very sensitive to flow phenomena. In particular, a steady reflow of unsaturated spins from outside the imaging plane results in high image intensities which depend on the flip angle, the repetition time, and the flow velocity. Here we describe a technique that suppresses these signals by saturating remote spins prior to entering the imaging plane, e.g., by intercalating the FLASH sequence with slice-selective 1-2-1 pulse packages. Experiments have been carried out on phantoms and human extremities. Flow artifacts in the phase-encoding direction of the images are strongly reduced. Vascular structures may be easily delineated using difference images obtained with and without flow suppression.

摘要

快速FLASH(快速低角度激发)核磁共振图像对流动现象非常敏感。特别是,不饱和自旋从成像平面外的稳定回流会导致高图像强度,这取决于翻转角、重复时间和流速。在此,我们描述一种技术,通过在自旋进入成像平面之前使其饱和来抑制这些信号,例如,通过在FLASH序列中插入切片选择的1-2-1脉冲包。已在模型和人体四肢上进行了实验。图像相位编码方向上的流动伪影大大减少。使用有和没有流动抑制时获得的差异图像,可以轻松勾勒出血管结构。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验