在 3T 扫描中比较 BLADE 和直线技术在肩部磁共振成像中的图像质量改善和运动伪影减少。

Image quality improvement and motion degradation reduction in shoulder MR imaging: comparison of BLADE and rectilinear techniques at 3-Tesla scanning.

机构信息

Department of Radiology, UT Southwestern Medical Center, Dallas, TX, 75022, USA.

Department of Orthopaedic Surgery, UT Southwestern Medical Center, Dallas, TX, USA.

出版信息

Skeletal Radiol. 2022 Dec;51(12):2291-2297. doi: 10.1007/s00256-022-04085-7. Epub 2022 Jun 25.

Abstract

OBJECTIVE

MR imaging of joints, particularly shoulder, requires a high degree of spatial resolution to ascertain anatomy and pathology. Unfortunately, motion artifacts can reduce the clinical quality of the examinations. BLADE sequence reduces motion degradation improving overall diagnostic imaging quality. The objective was to compare standard, rectilinear k-space coverage turbo spin echo (TSE), and BLADE sequences.

MATERIAL AND METHODS

Over a 4-month period, fifty-seven consecutive patients (22 males, 35 females; mean age: 48.5 years, range: 23-64 years) were scanned using traditional intermediate-weighted spin echo and BLADE sequences. Qualitative evaluation was performed by three musculoskeletal fellowship trained radiologists, each with more than 5 years of experience. Image sequences were evaluated using a Likert scale for each of the following five categories: motion degradation, ghosting/phase misregistration artifacts, star/radial encoding artifacts, fat suppression quality, and overall diagnostic quality. Additionally, image sequences were evaluated for signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) using manually drawn regions of interest (ROI) analysis.

RESULTS

Ghosting and phase artifacts were lower within BLADE sequence while streak artifacts were higher (p < 0.001). Image fat suppression, tendon and labral appearances, and the overall SNR and CNR were comparable on both sequences (p > 0.05).

CONCLUSION

Addition of BLADE reduces motion degradation and improves overall diagnostic imaging quality. Application of BLADE in patient scans suspected of motion artifacts can reduce the frequency of repeat imaging in patients with claustrophobia or those where motion is a concern. By reducing overall imaging time and call backs, it could reduce the cost burden to patients and healthcare system.

摘要

目的

关节的磁共振成像(MRI),尤其是肩部,需要高度的空间分辨率来确定解剖结构和病理变化。然而,运动伪影会降低检查的临床质量。BLADE 序列可减少运动伪影,从而提高整体诊断成像质量。本研究旨在比较标准的、直线型 K 空间覆盖的涡轮自旋回波(TSE)和 BLADE 序列。

材料与方法

在 4 个月的时间里,对 57 例连续患者(22 例男性,35 例女性;平均年龄:48.5 岁,范围:23-64 岁)进行了传统的中等权重自旋回波和 BLADE 序列扫描。由 3 名具有 5 年以上肌肉骨骼 fellowship培训经验的放射科医生进行了定性评估。使用李克特量表对以下五个类别中的每一个进行了图像序列评估:运动伪影、鬼影/相位配准伪影、星状/放射状编码伪影、脂肪抑制质量和整体诊断质量。此外,使用手动绘制的感兴趣区域(ROI)分析评估了图像序列的信噪比(SNR)和对比噪声比(CNR)。

结果

BLADE 序列中的鬼影和相位伪影较低,条纹伪影较高(p<0.001)。两种序列的图像脂肪抑制、肌腱和盂唇显示、整体 SNR 和 CNR 均相似(p>0.05)。

结论

添加 BLADE 可减少运动伪影并提高整体诊断成像质量。在怀疑有运动伪影的患者扫描中应用 BLADE,可以减少幽闭恐惧症或运动相关问题患者的重复成像频率。通过减少整体成像时间和召回,可以降低患者和医疗系统的成本负担。