Suppr超能文献

LAMax LAAC™ 左心耳封堵器在犬模型中的愈合反应:植入技术对愈合反应的潜在影响。

The healing response of LAMax LAAC™ left atrial appendage occluder in a canine model: the potential influence of the implantation technique on the healing response.

机构信息

Department of Cardiology, The Third Medical Center, Chinese PLA(People's Liberation Army) General Hospital, 69 Yongding Road, Hai Dian District, Beijing, 100039, China.

Faculty of Hepato-Pancreato-Biliary Surgery, Chinese PLA General Hospital, Beijing, China.

出版信息

BMC Cardiovasc Disord. 2022 Jun 27;22(1):291. doi: 10.1186/s12872-022-02731-5.

Abstract

BACKGROUND

Device-associated thrombus are potential causes for thromboembolic events post left atrial appendage closure (LAAC), and correlated with the complete endothelialization of the device surface. Our aim was to evaluate the endothelialization of LAMax LAAC™ occluder surface and analyze the potential influence of the implantation technique on the healing response.

METHODS

A total of 29 healthy dogs (28.0 ± 3.7 kg) were implanted with the devices successfully after ensuring COVER signs was met (Concavity of the disc, Oversizing by 20-50%, Verifying position, Ensuring stability, Residual flow < 5 mm by transesophageal echocardiographic (TEE) examination), and sacrificed at < 24 h, 1-, 2-, 3-, and 6-months. Gross examinations were conducted to evaluate healing response.

RESULTS

The mean diameters of LAA orifice measured by angiography and TEE were 19.0 ± 2.9 mm and 16.6 ± 2.9 mm (P < 0.05), respectively. TEE found that the discs in 18 dogs (62.1%) were completely pulled into the LAA with concavity and in 11 dogs incompletely pulled into the LAA with suboptimally concavity, while 5 of them had residual flow. Gross examinations showed that the complete endothelialization on the device surface with concaved disc was found at 1-month after LAAC. Microscopic examinations confirmed complete healing on the device with optimal closure effect.

CONCLUSIONS

The good healing response and the optimal closure effect were observed using the LAMax device in a canine model by following the COVER implantation technique.

摘要

背景

器械相关血栓是左心耳封堵(LAAC)术后发生血栓栓塞事件的潜在原因,与器械表面的完全内皮化有关。我们的目的是评估 LAMax LAAC™封堵器表面的内皮化,并分析植入技术对愈合反应的潜在影响。

方法

共 29 只健康犬(28.0±3.7kg)在满足 COVER 标志(盘的凹陷、放大 20-50%、验证位置、确保稳定、经食管超声心动图(TEE)检查残余血流<5mm)后成功植入设备,并在<24h、1 个月、2 个月、3 个月和 6 个月时处死。进行大体检查以评估愈合反应。

结果

造影和 TEE 测量的 LAA 口平均直径分别为 19.0±2.9mm 和 16.6±2.9mm(P<0.05)。TEE 发现 18 只狗(62.1%)的盘完全拉入 LAA 并呈凹陷状,11 只狗不完全拉入 LAA 且凹陷不理想,其中 5 只仍有残余血流。大体检查显示,LAAC 后 1 个月,盘呈凹陷状的器械表面完全内皮化。显微镜检查证实了具有最佳闭合效果的器械的完全愈合。

结论

通过遵循 COVER 植入技术,在犬模型中使用 LAMax 装置观察到良好的愈合反应和最佳的闭合效果。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/8425/9238264/3f9edc62274f/12872_2022_2731_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验