Yaminisharif Ahmad, Ahmadi Tafti Seyed Hossein, Hosseinsabet Ali, Shafiee Akbar
Tehran Heart Center, Cardiovascular Diseases Research Institute, Tehran University of Medical Sciences, Tehran, Iran.
J Innov Card Rhythm Manag. 2022 Jun 15;13(6):5057-5060. doi: 10.19102/icrm.2022.130604. eCollection 2022 Jun.
Lead-related complications compose a noticeable share of device-related complications. Pacemaker lead perforation is a recognized complication of lead implantation, particularly with active fixation leads, and should be considered in postoperative lead malfunction cases. We herewith present 3 challenging cases with ventricular pacemaker lead dispositioning who were successfully treated via percutaneous access.
与导线相关的并发症在与器械相关的并发症中占相当大的比例。起搏器导线穿孔是导线植入公认的并发症,尤其是主动固定导线,在术后导线故障病例中应予以考虑。我们在此介绍3例具有挑战性的心室起搏器导线处置病例,这些病例通过经皮途径成功治疗。