临床范围切伦科夫成像用于发现放射治疗质量改进机会:一年的经验。
One Year of Clinic-Wide Cherenkov Imaging for Discovery of Quality Improvement Opportunities in Radiation Therapy.
机构信息
Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire.
Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire; Dose Optics LLC, Lebanon, New Hampshire.
出版信息
Pract Radiat Oncol. 2023 Jan-Feb;13(1):71-81. doi: 10.1016/j.prro.2022.06.009. Epub 2022 Jun 29.
PURPOSE
Cherenkov imaging is clinically available as a radiation therapy treatment verification tool. The aim of this work was to discover the benefits of always-on Cherenkov imaging as a novel incident detection and quality improvement system through review of all imaging at our center.
METHODS AND MATERIALS
Multicamera Cherenkov imaging systems were permanently installed in 3 treatment bunkers, imaging continuously over a year. Images were acquired as part of normal treatment procedures and reviewed for potential treatment delivery anomalies.
RESULTS
In total, 622 unique patients were evaluated for this study. We identified 9 patients with treatment anomalies occurring over their course of treatment, which were only detected with Cherenkov imaging. Categorizing each event indicated issues arising in simulation, planning, pretreatment review, and treatment delivery, and none of the incidents were detected before this review by conventional measures. The incidents identified in this study included dose to unintended areas in planning, dose to unintended areas due to positioning at treatment, and nonideal bolus placement during setup.
CONCLUSIONS
Cherenkov imaging was shown to provide a unique method of detecting radiation therapy incidents that would have otherwise gone undetected. Although none of the events detected in this study reached the threshold of reporting, they identified opportunities for practice improvement and demonstrated added value of Cherenkov imaging in quality assurance programs.
目的
切伦科夫成像是一种临床可用的放射治疗验证工具。本研究的目的是通过对中心所有影像的回顾,发现始终开启的切伦科夫成像是一种新颖的事件检测和质量改进系统的优势。
方法和材料
多相机切伦科夫成像系统永久性安装在 3 个治疗掩体中,持续一年以上进行连续成像。图像是作为正常治疗程序的一部分获取的,并对潜在的治疗输送异常进行了审查。
结果
共有 622 名独特的患者接受了这项研究。我们发现了 9 名患者在治疗过程中出现了治疗异常,这些异常仅通过切伦科夫成像检测到。对每个事件进行分类表明,问题出现在模拟、计划、预处理审查和治疗输送中,而在这次审查之前,常规措施都没有检测到这些事件。在这项研究中确定的事件包括计划中剂量到达非预期区域、由于治疗时定位导致剂量到达非预期区域,以及在设置过程中不理想的敷贴放置。
结论
切伦科夫成像是一种独特的检测放射治疗事件的方法,否则这些事件将无法被检测到。尽管本研究中检测到的事件都没有达到报告的阈值,但它们为实践改进提供了机会,并证明了切伦科夫成像在质量保证计划中的附加价值。
相似文献
Int J Radiat Oncol Biol Phys. 2021-4-1
Tech Innov Patient Support Radiat Oncol. 2022-9-9
Radiother Oncol. 2021-7
J Med Imaging (Bellingham). 2020-5
引用本文的文献
Eur J Nucl Med Mol Imaging. 2025-5
Phys Med Biol. 2024-11-12
Int J Radiat Oncol Biol Phys. 2025-2-1
Molecules. 2024-6-19
本文引用的文献
Radiother Oncol. 2021-10
Radiother Oncol. 2021-7
Int J Radiat Oncol Biol Phys. 2021-4-1
Int J Radiat Oncol Biol Phys. 2020-1-1
Phys Med Biol. 2019-7-18