Department of Medical Imaging Technology, Manipal College of Health Professions, Manipal Academy of Higher Education, Manipal, India.
Department of Radio diagnosis and Imaging, Kasturba Medical College, Manipal Academy of Higher Education, Manipal, India.
J Xray Sci Technol. 2022;30(5):983-991. doi: 10.3233/XST-221172.
Pediatric population is more sensitive to the effects of radiation than adults. Establishing diagnostic reference level (DRL) is an efficient dose optimization technique implemented by many countries for reducing radiation dose during Computed Tomography (CT) examinations.
To estimate radiation dose and establish a new local diagnostic reference level for CT head examination in the pediatric population.
We prospectively recruited 143 pediatric patients referred for CT head examination with age ranging from 0-5 years old. All patients had undergone CT head examination using the standard pediatric head protocol. Volumetric CT dose index (CTDIvol) and dose length product (DLP) were recorded. The effective dose was first calculated. Then, 75th percentile of dose indices was calculated to establish DRLs.
DRLs in terms of CTDIvol and DLP are 23.84 mGy, 555.99 mGy.cm for patients <1 years old and 28.65 mGy, 794.99 mGy.cm for patients from 1-5 years old, respectively. Mean effective doses for <1 years old patients and 1-5 years old patients are 2.91 mSv and 2.78 mSv respectively.
The study concludes that DRL in terms of CTDIvol is lower but DRL in terms of DLP and the effective dose is higher compared to a few other studies which necessitate the need for dose optimization.
儿科人群对辐射的敏感性高于成年人。建立诊断参考水平(DRL)是许多国家实施的一种有效的剂量优化技术,可用于降低 CT 检查中的辐射剂量。
估计辐射剂量并为儿科人群的 CT 头部检查建立新的本地诊断参考水平。
我们前瞻性地招募了 143 名年龄在 0-5 岁的儿科患者,他们因 CT 头部检查而接受检查。所有患者均使用标准儿科头部协议进行 CT 头部检查。记录容积 CT 剂量指数(CTDIvol)和剂量长度乘积(DLP)。首先计算有效剂量,然后计算剂量指数的第 75 百分位数以建立 DRL。
对于<1 岁的患者,CTDIvol 和 DLP 的 DRL 分别为 23.84 mGy 和 555.99 mGy.cm,对于 1-5 岁的患者,CTDIvol 和 DLP 的 DRL 分别为 28.65 mGy 和 794.99 mGy.cm。<1 岁患者和 1-5 岁患者的平均有效剂量分别为 2.91 mSv 和 2.78 mSv。
与其他几项研究相比,本研究得出的 CTDIvol 方面的 DRL 较低,但 DLP 和有效剂量方面的 DRL 较高,这表明需要进一步进行剂量优化。