Davis L C, Radke G A
Anal Biochem. 1987 Feb 15;161(1):152-6. doi: 10.1016/0003-2697(87)90665-8.
We have used the bicinchoninic acid reagent developed by Pierce Chemical Co. to measure proteins in a simple flow injection analyzer. The sensitivity is comparable to that of the Lowry method and no pipetting of reagents is needed. Results are obtained in less than 1 min and samples may be run at a rate of 60/h. The response is linear over a range of protein concentration (0-10 micrograms) and sample size (5-20 microliters) convenient for most analytical requirements. A peristaltic pump, a controlled-temperature water bath, and a spectrophotometer with flow cuvette are the only special apparatus required.
我们使用了皮尔斯化学公司开发的二辛可宁酸试剂,在简单的流动注射分析仪中测量蛋白质。其灵敏度与洛瑞法相当,且无需移液操作。不到1分钟即可获得结果,样品运行速度可达每小时60个。在蛋白质浓度范围(0 - 10微克)和样品量(5 - 20微升)内响应呈线性,适合大多数分析需求。唯一需要的特殊仪器是蠕动泵、控温水浴和带流动比色皿的分光光度计。