Department of Neurosurgery, ABC Medical Center, Mexico City, Mexico.
Department of Anesthesiology, ABC Medical Center, Mexico City, Mexico.
Eur Spine J. 2022 Oct;31(10):2723-2732. doi: 10.1007/s00586-022-07297-6. Epub 2022 Jul 6.
We show a systematic review of known complications during intraoperative neuromonitoring (IONM) using transcranial electric stimulation motor evoked potentials (TES-MEP) on cervical spine surgery, which provides a summary of the main findings. A rare complication during this procedure, cardiac arrest by cardioinhibitory reflex, is also described.
Findings of 523 scientific papers published from 1995 onwards were reviewed in the following databases: CENTRAL, Cochrane Library, Embase, Google Scholar, Ovid, LILACS, PubMed, and Web of Science. This study evaluated only complications on cervical spine surgery undergoing TES-MEP IONM.
The review of the literature yielded 13 studies on the complications of TES-MEP IONM, from which three were excluded. Five studies are case series; the rest are case reports. Overall, 169 complications on 167 patients were reported in a total of 38,915 patients, a global prevalence of 0.43%. The most common complication was tongue-bite in 129 cases, (76.3% of all complication events). Tongue-bite had a prevalence of 0.33% (CI 95%, 0.28-0.39%) in all patients on TES-MEP IONM. A relatively low prevalence of severe complications was found: cardiac-arrhythmia, bradycardia and seizure, the prevalence of this complications represents only one case in all the sample. Alongside, we report the occurrence of cardiac arrest attributable to TES-MEP IONM.
This systematic review shows that TES-MEP is a safe procedure with a very low prevalence of complications. To our best knowledge, asystole is reported for the first time as a complication during TES-MEP IONM.
我们对使用经颅电刺激运动诱发电位(TES-MEP)进行颈椎手术时术中神经监测(IONM)的已知并发症进行了系统评价,总结了主要发现。还描述了该过程中一种罕见的并发症,即由心脏抑制反射引起的心脏骤停。
在以下数据库中对 1995 年以来发表的 523 篇科学论文的研究结果进行了综述:CENTRAL、Cochrane 图书馆、Embase、Google Scholar、Ovid、LILACS、PubMed 和 Web of Science。本研究仅评估了接受 TES-MEP IONM 的颈椎手术的并发症。
对 TES-MEP IONM 并发症的文献综述共产生了 13 项研究,其中 3 项被排除在外。五项研究为病例系列;其余的都是病例报告。共有 167 名患者的 169 例并发症在总计 38915 名患者中报告,全球患病率为 0.43%。最常见的并发症是 129 例舌咬伤(所有并发症事件的 76.3%)。在所有接受 TES-MEP IONM 的患者中,舌咬伤的患病率为 0.33%(CI 95%,0.28-0.39%)。发现相对较低的严重并发症患病率:心律失常、心动过缓和癫痫发作,所有样本中仅出现一例此类并发症。此外,我们报告了一例归因于 TES-MEP IONM 的心脏骤停。
本系统评价表明 TES-MEP 是一种安全的程序,并发症的患病率非常低。据我们所知,心脏停搏是 TES-MEP IONM 首次报告的并发症。