Helmet continuous positive airway pressure (CPAP) has been widely used during the COVID-19 pandemic. Specific filters (i.e. High Efficiency Particulate Air filter: HEPA; Heat & Moisture Exchanger Filter: HMEF) were used to prevent Sars-CoV2 environmental dispersion and were connected to the CPAP helmet. However, HEPA and HMEF filters may act as resistors to expiratory gas flow and increase the levels of pressure within the hood.
In a bench-top study, we investigated the levels of airway pressure generated by different HEPA and HMEF filters connected to the CPAP helmet in the absence of a Positive End Expiratory Pressure (PEEP) valve and with two levels of PEEP (5 and 10 cmHO). All steps were performed using 3 increasing levels of gas flow (60, 80, 100 L/min).
The use of 8 different commercially available filters significantly increased the pressure within the hood of the CPAP helmet with or without the use of PEEP valves. On average, the increase of pressure above the set PEEP ranged from 3 cmHO to 10 cmHO across gas flow rates of 60 to 100 L/min. The measure of airway pressure was highly correlated between the laboratory pressure transducer and the Helmet manometer. Bias with 95% Confidence Interval of Bias between the devices was 0.7 (-2.06; 0.66) cmHO.
The use of HEPA and HMEF filters placed before the PEEP valve at the expiratory port of the CPAP helmet significantly increase the levels of airway pressure compared to the set level of PEEP. The manometer can detect accurately the airway pressure in the presence of HEPA and HMEF filters in the helmet CPAP and its use should considered.
在 COVID-19 大流行期间,头盔持续气道正压通气(CPAP)已被广泛使用。特殊过滤器(即高效微粒空气过滤器:HEPA;热和湿气交换器过滤器:HMEF)用于防止 Sars-CoV2 环境扩散,并连接到 CPAP 头盔。然而,HEPA 和 HMEF 过滤器可能会对呼气气流产生阻力,并增加头盔内的压力水平。
在一项台式研究中,我们研究了在没有呼气末正压通气(PEEP)阀和使用两种 PEEP 水平(5 和 10 cmHO)的情况下,不同的 HEPA 和 HMEF 过滤器连接到 CPAP 头盔时产生的气道压力水平。所有步骤均使用 3 个增加的气体流量(60、80、100 L/min)进行。
使用 8 种不同的市售过滤器会显著增加 CPAP 头盔内的压力,无论是否使用 PEEP 阀。平均而言,在 60 至 100 L/min 的气体流量下,设置的 PEEP 以上的压力增加范围从 3 cmHO 到 10 cmHO。实验室压力传感器和头盔测压计之间的气道压力测量值高度相关。设备之间的偏差 95%置信区间的偏差为 0.7(-2.06;0.66)cmHO。
与设置的 PEEP 水平相比,将 HEPA 和 HMEF 过滤器放置在 CPAP 头盔呼气端口的 PEEP 阀之前,会显著增加气道压力水平。在头盔 CPAP 中有 HEPA 和 HMEF 过滤器的情况下,测压计可以准确地检测气道压力,应考虑使用。