Jongebloed W L, Worst J F
Doc Ophthalmol. 1986 Dec 30;64(1):143-52. doi: 10.1007/BF00166695.
A polypropylene surgical suture, which had served as fixation suture for an IOL for a period of 6.5 years, was removed from the eye of a Dutch patient because it had broken at one end and thus formed a potential risk for a corneal touch syndrome. After careful rinsing in 50% ethanol, to remove adhering debris, it was prepared for SEM. A virgin piece of polypropylene material from a surgical package underwent the same preparative procedure. The fixation suture showed cracks perpendicular to the longitudinal axis of the suture; part of the surface layer was nearly detached or completely missing; while the diameter of the suture was decreased towards both ends by over 50% in comparison with the original diameter. The exposed subsurface layer showed a fibrillar structure. The degradation phenomena are considered to be caused by the enzymatic action of tissue-fluids. Virgin material did not show any of the phenomena observed on the fixation suture under consideration.
一根曾作为人工晶状体固定缝线使用了6.5年的聚丙烯手术缝线,从一名荷兰患者眼中取出,因为其一端断裂,从而形成了角膜接触综合征的潜在风险。在50%乙醇中仔细冲洗以去除附着的碎屑后,将其制备用于扫描电子显微镜(SEM)观察。来自手术包装的一块未使用过的聚丙烯材料也经过相同的制备程序。固定缝线显示出垂直于缝线纵轴的裂缝;部分表层几乎分离或完全缺失;与原始直径相比,缝线两端的直径减小了超过50%。暴露的次表层呈现出纤维状结构。降解现象被认为是由组织液的酶促作用引起的。未使用过的材料未显示出在所考虑的固定缝线上观察到的任何现象。