Department of Hepatobiliary Surgery, The First Affiliated Hospital of Xi'an Jiaotong University, No. 277 West Yanta Road, Xi'an, Shaanxi, 710061, People's Republic of China.
Department of Thoracic Surgery, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, People's Republic of China.
Sci Rep. 2022 Jul 13;12(1):11916. doi: 10.1038/s41598-022-16050-4.
In single-port thoracoscopic lobectomy, surgical instruments are likely to collide and interfere with each other. We used magnetic anchor technique to design an anchoring device suitable for thoracoscopic surgery, and verified the safety and feasibility of its use in animal experiments. Ten Beagles were used as models, and magnetic anchor technology was used to assist thoracoscopic lobectomy. During the operation, a self-designed magnetic anchored internal grasper was used in place of the traditional laparoscopic grasping forceps. The operation time, intraoperative blood loss, incidence of postoperative complications, and the effect of intraoperative use of the device were analyzed. All 10 beagles were successfully operated; the mean operation time was 19.7 ± 3.53 min (range 15-26 min), and the postoperative blood loss was < 10 mL. No surgical complications occurred. During the operation, the internal grasper was firmly clamped, the auxiliary operation field was well exposed, and the interference of the main operation hole instruments was effectively reduced. We provide preliminary experimental evidence of the safety and feasibility of magnetic anchor technique-assisted thoracoscopic lobectomy.
在单孔胸腔镜肺叶切除术中,手术器械可能会相互碰撞和干扰。我们使用磁锚定技术设计了一种适用于胸腔镜手术的锚固装置,并通过动物实验验证了其使用的安全性和可行性。以 10 只比格犬为模型,采用磁锚定技术辅助胸腔镜肺叶切除术。在手术过程中,使用自行设计的磁锚定内置抓钳代替传统的腹腔镜抓钳。分析了手术时间、术中出血量、术后并发症发生率以及术中使用该装置的效果。10 只比格犬均成功完成手术,平均手术时间为 19.7±3.53 分钟(范围为 15-26 分钟),术后出血量<10 毫升。无手术并发症发生。手术过程中,内置抓钳夹持牢固,辅助手术野暴露良好,有效减少了主操作孔器械的干扰。我们为磁锚定技术辅助胸腔镜肺叶切除术的安全性和可行性提供了初步的实验证据。