Suppr超能文献

在 Elekta Versa HD 直线加速器中,评估 10MV 和 15MV 射束楔形和开放野的光中子剂量当量。

Evaluation of photoneutron dose equivalent in 10 MV and 15 MV beams for wedge and open fields in the Elekta Versa HD linac.

机构信息

Department of Medical Physics, College of Engineering, Anna University, Chennai, 600025, Tamilnadu, India.

Department of Medical Physics, College of Engineering, Anna University, Chennai, 600025, Tamilnadu, India.

出版信息

Appl Radiat Isot. 2022 Oct;188:110363. doi: 10.1016/j.apradiso.2022.110363. Epub 2022 Jul 7.

Abstract

In a high-energy medical linear accelerator (linac), if the interaction of photon energy is higher than the neutron binding energy of high atomic material, it emits a neutron field through a photonuclear reaction. The objective of this current study is to measure the photoneutron dose equivalent produces in a motorized wedge field and open field of 10 MV and 15 MV photon beams in Elekta Versa HD™ linac. The PNDE values were recorded at various positions along the patient plane using the Bubble Detector-Personal Neutron Dosimeter (BD-PND). The results revealed that the PNDE values are higher in 20 × 20 cm than 10 × 10 cm field sizes for both the 60° wedge and open fields of 10 MV and 15 MV beams. In addition, the 60° wedge fields generate higher photoneutron contamination when compared with the 45°, 30° wedge fields and open field sizes. Hence, on average PNDE values produced by the 15 MV beam were higher by a factor of 1.98 and 2.11 times for open and 60° wedge fields than the 10 MV beam, respectively.

摘要

在高能医用直线加速器(linac)中,如果光子能量的相互作用高于高原子材料的中子结合能,则会通过光核反应发射中子场。本研究的目的是测量 Elekta Versa HD™直线加速器中 10 MV 和 15 MV 光子束在电动楔形野和开放野中产生的光致中子剂量当量。使用气泡探测器-个人中子剂量计(BD-PND)在患者平面的各个位置记录 PNDE 值。结果表明,对于 60°楔形野和 10 MV 和 15 MV 射束的开放野,20×20 cm 的 PNDE 值高于 10×10 cm 的射野尺寸。此外,与 45°、30°楔形野和开放野尺寸相比,60°楔形野产生更高的光致中子污染。因此,平均而言,15 MV 射束产生的 PNDE 值分别比 10 MV 射束在开放野和 60°楔形野中高 1.98 倍和 2.11 倍。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验