[关于镫骨金属假体在核磁共振断层扫描仪磁场中行为的实验研究与理论思考]

[Experimental studies and theoretical considerations on the behavior of stapes metal prostheses in the magnetic field of a nuclear magnetic resonance tomograph].

作者信息

Hüttenbrink K B, Grosse-Nobis W

出版信息

Laryngol Rhinol Otol (Stuttg). 1987 Mar;66(3):127-30.

PMID:3586794
Abstract

The effects of Nuclear Magnetic Resonance Imaging fields on stapedectomy prostheses were studied in several experiments and calculated theoretically. Sufficiently strong forces to create a risk of dislodgement were found to be exerted on ferromagnetic alloys. However, with the paramagnetic steelwire prosthesis (Schuknecht) or the platinumpiston types no hazards are to be feared. Studies with operated patients proved the expected hearing loss inside the magnetic field at distinct positions of the head, as vibrations of the steelwire are attenuated.

摘要

在多项实验中研究了核磁共振成像场对镫骨切除术假体的影响,并进行了理论计算。结果发现,足够强的力会作用于铁磁合金,从而产生假体移位的风险。然而,对于顺磁性钢丝假体(舒克内希特型)或铂金活塞型假体,则无需担心有危害。对接受手术患者的研究证明,由于钢丝振动减弱,在头部不同位置的磁场中会出现预期的听力损失。