Wang Li, Goh Choon Heng
National Metrology Centre, Agency for Science, Technology and Research (A*STAR), Singapore, Singapore.
J Med Eng Technol. 2023 Jan;47(1):29-34. doi: 10.1080/03091902.2022.2096135. Epub 2022 Jul 27.
In view of the gaps between the current clinical thermometer standards and the commercially available products, in particular the introduction of forehead radiation thermometers into the market, ISO initiated a review on standard for clinical thermometers. The original clinical thermometer standard includes requirements for evaluating tympanic (ear) thermometers using identified reference blackbody sources. However, one of the challenges in reviewing the standard is to understand if the reference blackbody sources identified for ear thermometers are still usable for evaluating the laboratory accuracy of forehead thermometers. In order to answer this challenge, four brand new forehead thermometers from two different manufacturers are evaluated at NMC. Two reference blackbody sources, one JIS standard source with very small opening and one NMC source with larger opening, are used for the evaluation. In order to use JIS standard source, two types of adaptors, one supplied by one of the manufactures, one 3D printed, are evaluated. The measurement results together with their measurement uncertainties are presented and analysed. Suitable adaptors are identified and recommended.
鉴于当前临床体温计标准与市售产品之间存在差距,特别是额头辐射体温计进入市场,国际标准化组织(ISO)启动了对临床体温计标准的审查。最初的临床体温计标准包括使用已确定的参考黑体源评估鼓膜(耳部)体温计的要求。然而,审查该标准面临的挑战之一是要了解为耳部体温计确定的参考黑体源是否仍可用于评估额头体温计的实验室准确性。为应对这一挑战,在国家计量院对来自两家不同制造商的四台全新额头体温计进行了评估。使用了两种参考黑体源,一种是开口非常小的日本工业标准(JIS)源,另一种是开口较大的国家计量院源进行评估。为了使用JIS标准源,对两种类型的适配器进行了评估,一种由其中一家制造商提供,另一种是3D打印的。给出并分析了测量结果及其测量不确定度。确定并推荐了合适的适配器。