Stark Philipp Ch, Kalkbrenner Christoph, Klingler Werner, Brucher Rainer
Department of Mechatronics and Medical Engineering, University of Applied Science, Ulm, Germany.
Anaesthesiology, SRH Kliniken, Sigmaringen, Germany.
GMS Health Innov Technol. 2022 Jun 22;16:Doc03. doi: 10.3205/hta000135. eCollection 2022.
This paper presents a concept for detection of venous air emboli inside the superior vena cava using a central venous catheter with integrated Doppler ultrasound transducer installed on the tip. Several Doppler probes each with a single insonation frequencies of 2 MHz, 4 MHz or 8 MHz are characterized and compared for usefulness in this scenario. During in vitro experiments using an artificial blood circulatory with blood mimicking fluid bubbles with defined volumes were injected and recorded as gaseous embolic events. The in vitro results of measured embolus-blood-ratio values (EBR) in respect to the air bubbles volumes and its echogenicity showed a good correlation with the simulation model of spherical cross section scattering of such air bubbles. It is shown that the probe design still needs some improvements using a 4 MHz insonation frequency to get a useable detection sensitivity in such scenario within vena cava superior. The results suggest that it is possible to estimate the air bubble volume corresponding to the EBR using such a catheter probe.
本文提出了一种利用安装在尖端带有集成多普勒超声换能器的中心静脉导管检测上腔静脉内静脉空气栓塞的概念。对几个分别具有2兆赫、4兆赫或8兆赫单一超声频率的多普勒探头进行了特性分析,并比较了它们在这种情况下的实用性。在体外实验中,使用人工血液循环系统,注入模拟具有确定体积的液泡的血液,并将其记录为气态栓塞事件。关于气泡体积及其回声性的栓子-血液比率值(EBR)的体外测量结果与这种气泡的球形横截面散射模拟模型显示出良好的相关性。结果表明,在这种上腔静脉的情况下,使用4兆赫超声频率的探头设计仍需要一些改进,以获得可用的检测灵敏度。结果表明,使用这种导管探头可以估计与EBR相对应的气泡体积。