Plate, coil, and hollow-fiber cuprammonium cellulose dialyzers: discrepancy between incidence of anaphylactic reactions and degree of complement activation.

作者信息

Daugirdas J T, Potempa L D, Dinh N, Gandhi V C, Ivanovich P T, Ing T S

出版信息

Artif Organs. 1987 Apr;11(2):140-3. doi: 10.1111/j.1525-1594.1987.tb02646.x.

Abstract

During the past 10 years, the incidence of severe anaphylactic reactions during dialysis [type A first-use syndrome (FUS)] at our center has been much lower when using cuprammonium cellulose plate (CC-P) dialyzers (0/37, 750 dialyses) or coil (CC-C) dialyzers (0/32, 500) than when using cuprammonium cellulose hollow-fiber (CC-F) dialyzers (8/21,022 dialyses, p less than 0.005 by Chi-square). To determine if the difference in type A FUS incidence between the three dialyzer types could be explained by differences in complement activation, we compared plasma concentrations of C3a des-arginine (des arg) in patients undergoing dialysis with these three varieties of dialyzers. Plasma C3a des arg values increased markedly in the dialyzer outflow blood with the three dialyzer configurations. The levels were similar with the dialyzer types when results were corrected for membrane surface area. Also, the degree of leukopenia was not markedly different with the three dialyzer types. Our findings suggest that complement activation per unit surface area is similar during dialysis with plate, coil, and hollow-fiber cuprammonium cellulose dialyzers. The lack of correlation between the degree of complement activation and the incidence of type A FUS suggests that membrane-induced complement activation is not of primary importance to type A dialyzer hypersensitivity reactions.

摘要

在过去10年中,在我们中心,使用铜氨纤维素平板(CC-P)透析器(0/37,750次透析)或盘管(CC-C)透析器(0/32,500次)时,透析期间严重过敏反应(A型首次使用综合征,FUS)的发生率远低于使用铜氨纤维素中空纤维(CC-F)透析器时(8/21,022次透析,经卡方检验p<0.005)。为了确定三种透析器类型之间A型FUS发生率的差异是否可以用补体激活的差异来解释,我们比较了使用这三种不同透析器进行透析的患者血浆中C3a去精氨酸(des arg)的浓度。在三种透析器配置下,透析器流出血液中的血浆C3a des arg值均显著升高。当根据膜表面积校正结果时,不同透析器类型的水平相似。此外,三种透析器类型导致的白细胞减少程度没有明显差异。我们的研究结果表明,在使用平板、盘管和中空纤维铜氨纤维素透析器进行透析期间,单位表面积的补体激活情况相似。补体激活程度与A型FUS发生率之间缺乏相关性,这表明膜诱导的补体激活对A型透析器超敏反应并非至关重要。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验