Creare LLC, Hanover, New Hampshire 03755, USA.
Edare LLC, Lebanon, New Hampshire 03766, USA.
J Acoust Soc Am. 2022 Jul;152(1):601. doi: 10.1121/10.0012733.
This paper presents reference equivalent threshold sound pressure levels (RETSPLs) for the Wireless Automated Hearing Test System (WAHTS), a recently commercialized device developed for use as a boothless audiometer. Two initial studies were conducted following the ISO 389-9 standard [ISO 389-9 (2009). "Acoustics-Reference zero for the calibration of audiometric equipment. Part 9: Preferred test conditions for the determinations of reference hearing threshold levels" (International Organization for Standardization, Geneva)]. Although the standard recruitment criteria are intended to yield otologically normal test subjects, the recruited populations appeared to have slightly elevated thresholds [5-10 dB hearing level (HL)]. Comparison of WAHTS thresholds to other clinical audiometric equipment revealed bias errors that were consistent with the elevated thresholds of the RETSPL populations. As the objective of RETSPLs is to ensure consistent thresholds regardless of the equipment, this paper presents the RETSPLs initially obtained following ISO 389-9:2009 and suggested correction to account for the elevated HLs of the originally recruited populations. Two additional independent studies demonstrate the validity of these corrected thresholds.
本文提出了无线自动听力测试系统(WAHTS)的参考等效阈声压级(RETSPL),WAHTS 是一种最近商业化的设备,用于作为无耳罩听力计使用。该系统遵循 ISO 389-9 标准[ISO 389-9(2009)。“声学-听力计校准的参考零级。第 9 部分:参考听力阈值级别的测定的首选测试条件”(国际标准化组织,日内瓦)]进行了两项初始研究。尽管标准的招募标准旨在产生耳科正常的测试对象,但招募的人群似乎具有略微升高的阈值[5-10 dB 听力级(HL)]。WAHTS 阈值与其他临床听力计设备的比较显示出与 RETSPL 人群升高阈值一致的偏差误差。由于 RETSPL 的目的是确保无论设备如何都保持一致的阈值,因此本文最初根据 ISO 389-9:2009 提出了 RETSPL,并提出了纠正这些最初招募人群中 HLs 升高的建议。另外两项独立研究证明了这些校正阈值的有效性。