两种双能 X 射线吸收仪在儿童身体成分测量中的相互标定。

Cross-calibration of two dual-energy X-ray absorptiometry devices for the measurement of body composition in young children.

机构信息

Liggins Institute, The University of Auckland, Private Bag 92019, Auckland, New Zealand.

Department of Medicine and Department of General Practice and Primary Health Care, The University of Auckland, Auckland, New Zealand.

出版信息

Sci Rep. 2022 Aug 16;12(1):13862. doi: 10.1038/s41598-022-17711-0.

Abstract

This study aimed to cross-calibrate body composition measures from the GE Lunar Prodigy and GE Lunar iDXA in a cohort of young children. 28 children (mean age 3.4 years) were measured on the iDXA followed by the Prodigy. Prodigy scans were subsequently reanalysed using enCORE v17 enhanced analysis ("Prodigy enhanced"). Body composition parameters were compared across three evaluation methods (Prodigy, Prodigy enhanced, iDXA), and adjustment equations were developed. There were differences in the three evaluation methods for all body composition parameters. Body fat percentage (%BF) from the iDXA was approximately 1.5-fold greater than the Prodigy, whereas bone mineral density (BMD) was approximately 20% lower. Reanalysis of Prodigy scans with enhanced software attenuated these differences (%BF: - 5.2% [95% CI - 3.5, - 6.8]; and BMD: 1.0% [95% CI 0.0, 1.9]), although significant differences remained for all parameters except total body less head (TBLH) total mass and TBLH BMD, and some regional estimates. There were large differences between the Prodigy and iDXA, with these differences related both to scan resolution and software. Reanalysis of Prodigy scans with enhanced analysis resulted in body composition values much closer to those obtained on the iDXA, although differences remained. As manufacturers update models and software, researchers and clinicians need to be aware of the impact this may have on the longitudinal assessment of body composition, as results may not be comparable across devices and software versions.

摘要

本研究旨在对儿童队列中来自 GE Lunar Prodigy 和 GE Lunar iDXA 的身体成分测量值进行交叉校准。28 名儿童(平均年龄 3.4 岁)先在 iDXA 上进行测量,然后在 Prodigy 上进行测量。随后使用 enCORE v17 增强分析(“Prodigy enhanced”)重新分析 Prodigy 扫描。比较了三种评估方法(Prodigy、Prodigy enhanced、iDXA)下的身体成分参数,并制定了调整方程。三种评估方法的所有身体成分参数均存在差异。iDXA 的体脂百分比(%BF)比 Prodigy 高约 1.5 倍,而骨密度(BMD)低约 20%。使用增强软件重新分析 Prodigy 扫描减轻了这些差异(%BF:-5.2%[95%CI-3.5,-6.8];BMD:1.0%[95%CI0.0,1.9]),但除了 TBLH 总质量和 TBLH BMD 以及一些区域估计值外,所有参数仍存在显著差异。Prodigy 和 iDXA 之间存在较大差异,这些差异与扫描分辨率和软件都有关系。使用增强分析重新分析 Prodigy 扫描导致身体成分值更接近 iDXA 获得的值,尽管仍存在差异。随着制造商更新模型和软件,研究人员和临床医生需要意识到这可能对身体成分的纵向评估产生的影响,因为结果可能无法在不同设备和软件版本之间进行比较。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/6ff9/9381538/c89fbcae516f/41598_2022_17711_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验