Ma Hui, Chen Chun-Hua, Li Yan-Mei, Wang Jing-Jing, Hu Zhi-De
Department of Gastroenterology, the Affiliated Hospital of Inner Mongolia Medical University, Hohhot 010050, China.
Department of Laboratory Medicine, the Affiliated Hospital of Inner Mongolia Medical University, Hohhot 010050, China.
Clin Chim Acta. 2022 Oct 1;535:140-142. doi: 10.1016/j.cca.2022.08.011. Epub 2022 Aug 18.
Plasma D-dimer is a screening marker for thrombosis and hemostasis disorders. Falsely elevated D-dimer may result in misdiagnosis and potentially invasive investigations. Therefore, identifying falsely elevated D-dimer is of great value in laboratory practice. D-dimer is determined by immunoassay and vulnerable to heterophilic antibody interference. Here, we reported a case of falsely elevated D-dimer partially caused by heterophilic antibodies. Dilution test, polyethylene glycol precipitation, heterophilic blocking reagent and method comparison were used to identify the heterophilic antibody interference. All these methods, except for the heterophilic blocking reagent, revealed the existence of heterophilic antibody interference. Polyethylene glycol precipitation failed to reduce the D-dimer to its reference interval. Therefore, we conclude that the falsely elevated D-dimer is partially caused by heterophilic antibody interference.
血浆D-二聚体是血栓形成和止血障碍的筛查标志物。D-二聚体假性升高可能导致误诊以及潜在的侵入性检查。因此,在实验室实践中识别D-二聚体假性升高具有重要价值。D-二聚体通过免疫测定法测定,易受嗜异性抗体干扰。在此,我们报告了1例部分由嗜异性抗体导致D-二聚体假性升高的病例。采用稀释试验、聚乙二醇沉淀法、嗜异性阻断试剂和方法比较来识别嗜异性抗体干扰。除嗜异性阻断试剂外,所有这些方法均显示存在嗜异性抗体干扰。聚乙二醇沉淀法未能将D-二聚体降至参考区间。因此,我们得出结论,D-二聚体假性升高部分是由嗜异性抗体干扰引起的。