Suppr超能文献

二硫苏糖醇用于纠正D-二聚体假性升高的应用及浓度选择

Application and concentration selection of dithiothreitol for correcting false elevation of D-dimer.

作者信息

Zeng Jia-Wei, Song Guang-Hao, Xia Mao

机构信息

Medical Laboratory of Taizhou Fourth People's Hospital, Taizhou, Jiangsu, 225300, China.

Department of Laboratory Medicine, The Affiliated Drum Tower Hospital, Medical School of Nanjing University, Nanjing, Jiangsu, 210008, China.

出版信息

Pract Lab Med. 2025 Mar 20;45:e00469. doi: 10.1016/j.plabm.2025.e00469. eCollection 2025 Jul.

Abstract

BACKGROUND

Immunological methods for D-dimer detection are prone to interference, leading to falsely elevated values. This study evaluates the optimal concentration and corrective efficacy of dithiothreitol (DTT) in addressing false D-dimer elevations.

METHODS

Plasma samples from seven patients with falsely elevated D-dimer levels, identified using the Sysmex CS5100 coagulation analyzer, were analyzed. Correction was performed using the Stago STA-R Max analyzer with alternative detection antibodies. Additionally, plasma samples from thirty patients with confirmed D-dimer elevations were treated with varying concentrations of DTT (0.01, 0.05, 0.10, and 0.20 mol/L) and normal saline (NS). The optimal concentration of 0.01 mol/L DTT was identified and applied to the experimental group. The results were compared with those obtained using an alternative immunoassay method.

RESULTS

In the true elevation group, samples treated with 0.01 mol/L DTT showed no significant difference in D-dimer levels compared to the NS control or original results (all P > 0.05). Higher concentrations of DTT significantly lowered D-dimer levels compared to the NS group (all P < 0.05). In the experimental group, 0.01 mol/L DTT treatment and the alternative immunoassay method similarly reduced D-dimer levels, with no significant difference between the two correction methods (P > 0.05).

CONCLUSION

A 0.01 mol/L DTT solution effectively corrects falsely elevated D-dimer levels without affecting true elevations, yielding results comparable to those achieved by changing the immunoassay method.

摘要

背景

D - 二聚体检测的免疫方法容易受到干扰,导致数值假性升高。本研究评估二硫苏糖醇(DTT)在解决D - 二聚体假性升高方面的最佳浓度和校正效果。

方法

分析了使用Sysmex CS5100凝血分析仪鉴定出的7例D - 二聚体水平假性升高患者的血浆样本。使用带有替代检测抗体的Stago STA - R Max分析仪进行校正。此外,对30例确诊D - 二聚体升高的患者的血浆样本用不同浓度的DTT(0.01、0.05、0.10和0.20 mol/L)和生理盐水(NS)进行处理。确定了0.01 mol/L DTT的最佳浓度并应用于实验组。将结果与使用替代免疫测定方法获得的结果进行比较。

结果

在真性升高组中,用0.01 mol/L DTT处理的样本与NS对照组或原始结果相比,D - 二聚体水平无显著差异(所有P > 0.05)。与NS组相比,较高浓度的DTT显著降低了D - 二聚体水平(所有P < 0.05)。在实验组中,0.01 mol/L DTT处理和替代免疫测定方法同样降低了D - 二聚体水平,两种校正方法之间无显著差异(P > 0.05)。

结论

0.01 mol/L DTT溶液可有效校正假性升高的D - 二聚体水平,而不影响真性升高,其结果与改变免疫测定方法获得的结果相当。