Dose intercomparison at the radiotherapy centres in The Netherlands. 1. Photon beams under reference conditions and for prostatic cancer treatment.

作者信息

Wittkämper F W, Mijnheer B J, van Kleffens H J

出版信息

Radiother Oncol. 1987 May;9(1):33-44. doi: 10.1016/s0167-8140(87)80217-7.

Abstract

In 1985, a dosimetry intercomparison was performed at all 20 radiotherapy centres in The Netherlands. Absorbed dose was determined with an ionization chamber under reference conditions in a water phantom for cobalt-60 gamma-ray and megavoltage X-ray beams. The mean difference between measured and stated dose values was 0.5% with a standard deviation of 1.9%, but up to 6% at maximum. As soon as all institutes apply a common dosimetry protocol, this maximum difference will reduce to about 2%. In addition, an anthropomorphic phantom was irradiated to simulate the treatment of a prostatic cancer. The dose, determined with an ionization chamber at the isocentre and thermoluminescent dosimeters (TLD) powder at several points situated in the target volume, the bladder and the rectum, was compared with the stated dose calculated with the local planning system. Only small differences were found between the measured and stated dose at the isocentre: on the average 1.5%, with a standard deviation of 1.5%. The difference between stated and measured dose at several points situated in the target volume was on the average 0.4%, with a standard deviation of 5.2%. Almost the same result was found for a point situated in the bladder. In the rectum, the average difference was about 4%, however, with a large standard deviation, 18%, due to the relatively steep dose gradient at these points.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验