中欧质量保证网络。光子束输出校准的外部审计。

Quality assurance network in central Europe. External audit on output calibration for photon beams.

作者信息

Izewska J, Novotny J, Gwiazdowska B, Kindlova A, Kontra G, van Dam J, Dutreix A, van der Schueren E

机构信息

Radiotherapy Department, University Hospital Gasthuisberg, Leuven, Belgium.

出版信息

Acta Oncol. 1995;34(6):829-38. doi: 10.3109/02841869509127193.

Abstract

The EROPAQ project for TLD monitoring of photon beams started in June 1994 with the set-up of the TLD system: calibration, reading and evaluation procedures. The acceptance level of +/- 3% was set for the TLD intercomparisons. The policy of the project was to check all beams in 47 participating radiotherapy centres and to recheck all the beams in those centres, where a deviation exceeding +/- 3% occurred in one or more of the beams. Out of 129 beams checked, 100 beams (78%) were found within the +/- 3% limit. Eleven beams show deviations larger than +/- 6%, and immediate corrective action was undertaken. Out of 47 centres checked, 22 did not participate in any external audit in a preceding 5 years. In these centres 68% (34/50) of the total number of gamma and x-ray beams checked but only 59% (20/34) of gamma beams were within the acceptance level, while in the 25 centres, which participated in an external audit before, these figures were 84% (66/79) and 88% (35/40) respectively. The sources of discrepancies were thoroughly investigated, discussed with the participants and the errors corrected. Poor results were in several cases associated with very old design of radiotherapy units and old dosimetry systems, equipped with inadequate ionization chambers. In several centres, an insufficient training of the physicists in clinical dosimetry was observed. Thanks to the corrective action, a great improvement of calibration of the beams was achieved. Standard deviation of the distribution of the results for all x and gamma beams checked decreased from SD = 7.4% at the first check to SD = 2.5% at the second check.

摘要

用于光子束TLD监测的EROPAQ项目始于1994年6月,当时建立了TLD系统:校准、读数和评估程序。TLD比对的验收水平设定为±3%。该项目的政策是检查47个参与放疗中心的所有光束,并对那些有一束或多束光束偏差超过±3%的中心的所有光束进行重新检查。在检查的129束光束中,发现100束(78%)在±3%的限值内。11束光束显示偏差大于±6%,并立即采取了纠正措施。在检查的47个中心中,有22个在过去5年中未参与任何外部审核。在这些中心,检查的伽马射线和X射线光束总数的68%(34/50)在验收水平内,但伽马光束只有59%(20/34)在验收水平内,而在之前参与过外部审核的25个中心,这些数字分别为84%(66/79)和88%(35/40)。对差异来源进行了深入调查,与参与者进行了讨论,并纠正了错误。在一些情况下,结果不佳与放疗设备的非常陈旧的设计以及配备不足电离室的旧剂量测定系统有关。在几个中心,观察到物理学家在临床剂量测定方面的培训不足。由于采取了纠正措施,光束校准有了很大改进。所有检查的X射线和伽马射线光束结果分布的标准差从第一次检查时的SD = 7.4%降至第二次检查时的SD = 2.5%。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验