A dosimetric intercomparison of megavoltage photon beams in UK radiotherapy centres.

作者信息

Thwaites D I, Williams J R, Aird E G, Klevenhagen S C, Williams P C

机构信息

Department of Medical Physics and Medical Engineering, University of Edinburg, UK.

出版信息

Phys Med Biol. 1992 Feb;37(2):445-61. doi: 10.1088/0031-9155/37/2/011.

Abstract

A dosimetry intercomparison has been carried out for all 64 radiotherapy centres in the UK. Doses were measured with an ionization chamber in an epoxy resin water-substitute phantom of relatively simple geometry. Reference-point measurements were made for all MV photon beams. For 61 Co-60 beams, a mean ratio of measured-to-stated dose of 1.002 was observed with a standard deviation of 0.014, whilst for 100 MV x-ray beams, the corresponding figures were 1.003 and 0.015. 97% of beams lay within a +/- 3% deviation. One measurement was instrumental in discovering a large discrepancy. Doses were also investigated in two planned three-field distributions at one beam quality in each centre. One of these was in a homogeneous phantom, whilst the second included a lung-equivalent insert. Doses were measured at the central point and at four other points in the high dose volume. In both situations, the mean ratio of measured-to-calculated doses for all points was 1.008, with standard deviations of 0.027 and 0.035 for the uniform and non-uniform phantoms, respectively. Discrepancies over 5% were followed up. The work must be viewed in the context of other international intercomparisons and is an essential part of wider radiotherapy audit processes.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验