ENT Unit, Department of Surgical Sciences, University of Turin, Turin.
Department of Public Health Sciences and Paediatrics, University of Torino, Torino, Italy.
Otol Neurotol. 2022 Oct 1;43(9):1000-1004. doi: 10.1097/MAO.0000000000003671. Epub 2022 Aug 30.
To evaluate speech in noise results and subjective benefit in bilateral active bone conduction implant (ABCI) for bilateral mixed hearing loss.
Prospective, comparative.
Ear, Nose and Throat Unit, Department of Surgical Sciences, University of Turin.
Seven patients with conductive/mixed hearing loss.
Patients underwent simultaneous or sequential bilateral surgery for ABCI.
The speech intelligibility in noise was assessed with the Ita Matrix test in summation, squelch and head shadow settings. First, the tests were performed with one device activated in the ear with lower speech recognition score, then with both devices. Patients filled in an Abbreviated Profile of Hearing Aid Benefit (APHAB) questionnaire to investigate the hearing aid benefit.
When bilateral devices were activated, an improvement of signal-to-noise ratio was observed in all settings. The difference between bilateral and unilateral stimulation is 4.66 dB ( p = 0.016) in the summation, 2.24 dB ( p = 0.047) in the squelch, 7.50 dB ( p = 0.016) in the head shadow setting.Looking at the APHAB global score (GS), patients report lower mean scores, hence less difficulties, when using two devices (GS, 21.9%; standard deviation (SD), 8.28) rather than one (GS, 33.0%; SD, 10.24) ( p = 0.018).
In symmetric mixed bilateral hearing loss, rehabilitation with an ABCI fitted bilaterally shows audiologic advantages in speech perception in noise, not only thanks to the summation effect and by reducing head shadow but also by improving the binaural unmasking based on the squelch effect. Audiometric outcomes are confirmed by the GSs obtained in the APHAB questionnaire.
评估双侧主动骨导植入物(ABCI)治疗双侧混合性听力损失的语音在噪声中的表现和主观获益。
前瞻性、对照研究。
都灵大学外耳、鼻喉科。
7 例传导性/混合性听力损失患者。
患者接受双侧 ABCI 的同期或序贯双侧手术。
采用 Ita 矩阵测试在总和、消声和头影设置下评估语音可懂度。首先,在听力识别得分较低的耳朵中激活一个设备进行测试,然后再激活两个设备进行测试。患者填写简短听力助听效果问卷(APHAB)以调查听力助听效果。
当双侧设备激活时,在所有设置中均观察到信噪比提高。双侧刺激与单侧刺激之间的差异在总和设置中为 4.66dB(p=0.016),在消声设置中为 2.24dB(p=0.047),在头影设置中为 7.50dB(p=0.016)。从 APHAB 总体评分(GS)来看,当使用两个设备时(GS,21.9%;标准差(SD),8.28),患者报告的平均评分较低,即使用两个设备时的困难较小(GS,33.0%;SD,10.24)(p=0.018)。
在对称混合性双侧听力损失中,双侧配备 ABCI 的康复在噪声中的语音感知方面具有听力优势,这不仅得益于总和效应和减少头影,还得益于基于消声效应改善双耳掩蔽。听力计结果通过 APHAB 问卷中的 GS 得到证实。