Suppr超能文献

使用剂量平衡函数计算体部 CT 扫描纵向位置的扫描长度和剂量面积乘积。

Calculation of scan length and size-specific dose at longitudinal positions of body CT scans using dose equilibrium function.

机构信息

Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts, USA.

出版信息

Med Phys. 2023 Feb;50(2):1096-1104. doi: 10.1002/mp.15959. Epub 2022 Sep 7.

Abstract

BACKGROUND

Dose evaluation at longitudinal positions of body computed tomography (CT) scans is useful for CT quality assurance programs and patient organ dose evaluation. Accurate estimates depend on both patient size and scan length.

PURPOSE

To propose practical evaluation of the average dose to the transverse slab of an axial image slice for adult body CT examinations, considering not only patient size but also scan length, and to compare the results with those of Monte Carlo (Geant4) simulation [D (z)] and size-specific dose estimates at longitudinal positions of scans [SSDE(z)] from international standards (IEC publication no. 62985).

METHODS

In a scan series, the total dose at each z-axis location was calculated using the input information identical to the SSDE(z) evaluation. Each axial image slice (slice thickness, 2.5 or 5 mm) was first considered independently. Its z-axis coverage and CTDI (from the DICOM headers) were used to directly calculate a z-axis dose profile for the average dose over the cross-section of a water phantom, using the approach to equilibrium function. The phantom diameter was taken to be equal to the patient water equivalent diameter at that slice. The above was repeated at all slices and the dose at each z-axis location was accumulated from all profiles, referred to as D (z). For validation, we considered a cohort of 65 patients, who underwent chest and abdominopelvic examinations. The resultant D (z) was compared with D (z) and SSDE(z), both available in a previous paper.

RESULTS

D (z) evaluation could be used to accurately assess the scan range average dose, with an accuracy of 7.1%-8.7% for 65 patients in two examinations. On individual image slices, the maximum difference in magnitude between D (z) and D (z) [and between SSDE(z) and D (z) in parentheses] was 37.5% (85%) [two edges (2 × 5 cm) of chest scan range], 17.8% (35.2%) (the remaining central region of chest scan), 26.8% (74.1%) [two edges (2 × 5 cm) of abdominopelvic scan range], and 14.2% (22.5%) (the remaining central region of abdominopelvic scan).

CONCLUSIONS

Identical input data are used for D (z) and SSDE(z) evaluations. The latter is limited to the z-axis locations within scan range. At each image slice, SSDE(z) is equivalent to the midpoint dose of a fixed-mA scan of 15-30 cm (scan length). In contrast, D (z) considers dose accumulation from varying scan length (from sub-centimeter to about 1 m) and tube current, and dose profile is also computed outside scan range. Besides greatly improving dose evaluation for individual image slices, D (z) allows for evaluating dose accumulation from multiple series, which typically span different scan ranges. Our proposal may assist CT manufacturers and dose index monitoring software in assessing dose at longitudinal positions of body CT scans.

摘要

背景

在体部计算机断层扫描(CT)的纵向位置进行剂量评估对于 CT 质量保证计划和患者器官剂量评估非常有用。准确的估计取决于患者的大小和扫描的长度。

目的

提出一种用于评估成人身体 CT 检查的轴向图像切片横向平面平均剂量的实用方法,不仅考虑患者的大小,还考虑扫描的长度,并将结果与国际标准(IEC 出版物第 62985 号)中的蒙特卡罗(Geant4)模拟[D(z)]和扫描纵向位置的特定尺寸剂量估计(SSDE(z))进行比较。

方法

在扫描系列中,使用与 SSDE(z)评估相同的输入信息计算每个 z 轴位置的总剂量。首先独立考虑每个轴向图像切片(切片厚度为 2.5 或 5 毫米)。它的 z 轴覆盖范围和 CTDI(来自 DICOM 标头)用于直接计算水模体横截面的 z 轴剂量分布,使用平衡函数方法。将体模直径取为等于该切片处的患者水等效直径。在所有切片上重复上述步骤,并从所有分布中累积每个 z 轴位置的剂量,称为 D(z)。为了验证,我们考虑了一组 65 名接受胸部和腹部盆腔检查的患者。将得到的 D(z)与之前一篇论文中提供的 D(z)和 SSDE(z)进行比较。

结果

D(z)评估可用于准确评估扫描范围的平均剂量,对于两个检查中的 65 名患者,精度为 7.1%-8.7%。在个别图像切片上,D(z)和 D(z)之间的最大幅度差异[括号内为 SSDE(z)和 D(z)之间的差异]为 37.5%(85%)(胸部扫描范围的两个边缘(2×5 厘米)),17.8%(35.2%)(胸部扫描的剩余中央区域),26.8%(74.1%)(腹部盆腔扫描范围的两个边缘)和 14.2%(22.5%)(腹部盆腔扫描的剩余中央区域)。

结论

D(z)和 SSDE(z)评估使用相同的输入数据。后者仅限于扫描范围内的 z 轴位置。在每个图像切片上,SSDE(z)等效于 15-30cm(扫描长度)的固定毫安扫描的中点剂量。相比之下,D(z)考虑了来自不同扫描长度(从亚厘米到约 1 米)和管电流的剂量累积,并且还在扫描范围之外计算剂量分布。除了大大提高对个别图像切片的剂量评估外,D(z)还允许评估来自多个系列的剂量累积,这些系列通常跨越不同的扫描范围。我们的建议可以帮助 CT 制造商和剂量指数监测软件评估体部 CT 扫描的纵向位置的剂量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验