Mjöberg B, Rydholm A, Selvik G, Onnerfält R
Acta Orthop Scand. 1987 Apr;58(2):106-8. doi: 10.3109/17453678709146450.
Sixteen patients were examined by roentgen stereophotogrammetry during the first year after total hip arthroplasty. Eight prostheses were fixed with low-viscosity and eight with high-viscosity cement. Eight acetabular components, four in each group, migrated cranially, and three femoral components, all in the low-viscosity group, migrated distally. Our findings suggest that low-viscosity cement does not improve prosthetic fixation.